Maino Paulo, Dullenkopf Alexander, Keller Christian, Bernet-Buettiker Vera, Weiss Markus
Department of Anaesthesia, University Children's Hospital, Zurich, Switzerland.
Paediatr Anaesth. 2006 Jan;16(1):25-30. doi: 10.1111/j.1460-9592.2005.01672.x.
Hyperinflation of LMA cuffs carries the risk of airway morbidity by exerting pressure on laryngeal and pharyngeal structures. Cuff hyperinflation in LMAs can result from nitrous oxide diffusion into the LMA cuff and from deliberate manual cuff inflation.
In an in vitro set up, maximum recommended cuff filling volumes for size 1, 1.5, 2, 2.5, 3 disposable LMAs from different manufacturers [SoftSeal LMA (Portex); Unique LMA (Intavent); Marshall LAD, LaryngoSeal (Tyco)] and reusable Classic and ProSeal LMAs (Intavent) were inflated into completely emptied LMA-cuffs and into LMA-cuffs at resting volume. Cuff pressures were measured using a cuff manometer. Experiments were performed eight times using two exemplars of each brand/size.
Maximum recommended cuff filling volumes for pediatric LMAs resulted in hyperinflation (cuff pressure >60 cmH(2)O) in almost all LMAs, starting with emptied cuff. Starting from resting cuff volume, the maximum recommended inflation volume resulted in cuff pressures of >120 cmH(2)O in all LMAs and sizes except in the ProSeal size 3 (101 +/- 1 cmH(2)O).
Since the volume of air which is effectively required depends on several factors and varies between patients, the cuffs should be inflated only with the minimum volume of air required to form an effective seal with the respiratory and gastrointestinal tracts and the cuff pressure should be controlled.
喉罩(LMA)套囊过度充气会对喉和咽结构施加压力,从而带来气道发病风险。LMA套囊过度充气可能是由于氧化亚氮扩散至LMA套囊以及故意手动给套囊充气所致。
在体外实验中,将来自不同制造商的1号、1.5号、2号、2.5号、3号一次性LMA[SoftSeal LMA(Portex公司);Unique LMA(Intavent公司);Marshall LAD、LaryngoSeal(泰科公司)]以及可重复使用的Classic和ProSeal LMA(Intavent公司)的最大推荐套囊填充量,分别充入完全排空的LMA套囊以及处于静息容量的LMA套囊中。使用套囊压力计测量套囊压力。每个品牌/尺寸的两个样本进行八次实验。
儿科LMA的最大推荐套囊填充量几乎在所有LMA中,从排空的套囊开始就导致了套囊过度充气(套囊压力>60 cmH₂O)。从套囊静息容量开始,除了ProSeal 3号尺寸(101±1 cmH₂O)外,所有LMA和尺寸的最大推荐充气量都导致套囊压力>120 cmH₂O。
由于有效所需的空气量取决于多种因素且在患者之间存在差异,因此套囊应仅用形成与呼吸道和胃肠道有效密封所需的最小空气量进行充气,并且应控制套囊压力。