Jamali Amir A, Lozynsky Andrew J, Harris William H
Department of Orthopaedic Surgery, UC Davis Medical Center, Sacramento, California, USA.
J Arthroplasty. 2006 Jan;21(1):122-8. doi: 10.1016/j.arth.2005.05.018.
The search for improved femoral fixation in cemented total hip arthroplasty is ongoing. Two design variables, surface finish and stem contour, were evaluated. Sixteen titanium femoral stems of one design were cemented into fiberglass femora. One half of the components had a polished surface and the rest had a roughened finish. Within each group, 4 stems had vertically oriented ribs on the proximal portion and 4 did not. Micromotion was measured in a stair climbing simulator with loading to a joint reaction force of 200 kg for 6 million cycles. Micromotion increased throughout the course of the experiment. Stems with a polished surface had significantly higher micromotion. Although stems with ribs had less micromotion compared with those without ribs, this difference was not statistically significant.
在骨水泥型全髋关节置换术中,对改进股骨固定的研究仍在进行。评估了两个设计变量,即表面光洁度和柄部轮廓。将16个一种设计的钛制股骨干柄骨水泥固定到玻璃纤维股骨中。一半的组件表面抛光,其余的表面粗糙。在每组中,4个股骨干柄近端有垂直方向的肋条,4个没有。在楼梯攀爬模拟器中测量微动,加载至关节反应力200 kg,持续600万次循环。在整个实验过程中,微动增加。表面抛光的股骨干柄微动明显更高。虽然有肋条的股骨干柄比没有肋条的微动少,但这种差异无统计学意义。