Mazzocchi S, Belli G, Busoni S, Gori C, Menchi I, Salucci P, Taddeucci A, Zatelli G
S.O.D. Fisica Sanitaria, Azienda Ospedaliero-Universitaria Careggi, Viale Pieraccini 17, Florence, Italy.
Radiat Prot Dosimetry. 2005;117(1-3):169-73. doi: 10.1093/rpd/nci743. Epub 2006 Feb 3.
Phototimer set-up is a critical procedure for dose and image quality optimisation in computed radiography (CR) systems. While a conventional radiography automatic exposure control device (AEC) can be calibrated in order to gain a constant optical density on the film independent of beam quality and patient size, CR detectors present a high dynamic range which allows a much larger dose interval, but with different image quality levels. CR leads to a less frequent exam repetition, but may produce quite noisy images if the exposure level on the plate is not correct. The aim of this work is to evaluate the performance of a CR plate (Agfa MD40) in order to optimally calibrate an AEC device. The plate response has been characterised in terms of digital signal, exposure on the plate and signal-to-noise ratio for different beam qualities, in a patient of standard size.
光定时器设置是计算机X线摄影(CR)系统中剂量和图像质量优化的关键程序。虽然传统的X线摄影自动曝光控制设备(AEC)可以进行校准,以便在胶片上获得与射线质和患者体型无关的恒定光学密度,但CR探测器具有高动态范围,这允许更大的剂量间隔,但图像质量水平不同。CR导致检查重复频率较低,但如果平板上的曝光水平不正确,可能会产生噪声较大的图像。这项工作的目的是评估CR平板(爱克发MD40)的性能,以便对AEC设备进行最佳校准。在标准体型的患者中,已根据数字信号、平板上的曝光以及不同射线质下的信噪比来表征平板响应。