Hsieh H Ben, Marrinucci Dena, Bethel Kelly, Curry Douglas N, Humphrey Mark, Krivacic Robert T, Kroener Joan, Kroener Lindsay, Ladanyi Andras, Lazarus Nicole, Kuhn Peter, Bruce Richard H, Nieva Jorge
The Scripps-PARC Institute for Advanced Biomedical Science, Palo Alto Research Center, 3333 Coyote Hill Rd., Palo Alto, CA 94304, United States.
Biosens Bioelectron. 2006 Apr 15;21(10):1893-9. doi: 10.1016/j.bios.2005.12.024. Epub 2006 Feb 7.
Epithelial tumor cells circulate in peripheral blood at ultra-low concentrations in cancer patients. We have developed an instrument capable of rapid and accurate detection of rare cells in circulation utilizing fiber-optic array scanning technology (FAST). The FAST cytometer can locate immunofluorescently labeled rare cells on glass substrates at scan rates 500 times faster than conventional automated digital microscopy. These high scan rates are achieved by collecting fluorescent emissions using a fiber bundle with a large (50 mm) field of view. Very high scan rates make possible the ability to detect rare events without the requirement for an enrichment step. The FAST cytometer was used to detect, image and re-image circulating tumor cells in peripheral blood of breast cancer patients. This technology has the potential to serve as a clinically useful point-of-care diagnostic and a prognostic tool for cancer clinicians. The use of a fixed substrate permits the re-identification and re-staining of cells allowing for additional morphologic and biologic information to be obtained from previously collected and identified cells.
上皮肿瘤细胞在癌症患者外周血中以超低浓度循环。我们开发了一种利用光纤阵列扫描技术(FAST)能够快速准确检测循环中稀有细胞的仪器。FAST细胞仪能够以比传统自动数字显微镜快500倍的扫描速度在玻璃基板上定位免疫荧光标记的稀有细胞。这些高扫描速度是通过使用具有大视野(50毫米)的光纤束收集荧光发射来实现的。非常高的扫描速度使得无需富集步骤就能检测稀有事件成为可能。FAST细胞仪用于检测、成像和重新成像乳腺癌患者外周血中的循环肿瘤细胞。这项技术有潜力作为一种对癌症临床医生有用的床旁诊断和预后工具。使用固定基板允许对细胞进行重新识别和重新染色,从而从先前收集和识别的细胞中获得额外的形态学和生物学信息。