A 2-year retrospective analysis of laboratory testing for activated protein C resistance with a factor V-corrected activated partial thromboplastin time-based method.

作者信息

De Bel Annelies V, Van der Cruyssen Greta A, Devreese Katrien M

机构信息

Laboratory of Clinical Biology, Ghent University Hospital, Ghent, Belgium.

出版信息

Blood Coagul Fibrinolysis. 2006 Mar;17(2):155-60. doi: 10.1097/01.mbc.0000203861.47492.db.

Abstract

The factor V-corrected activated protein C resistance assay is the test of choice to screen for the factor V Leiden mutation. During the past 2 years, local test results with the frequently used Coatest APCR kit were evaluated and compared with the results of DNA analysis, the 'gold standard'. Samples of 278 patients were analysed by both techniques. We were unable to confirm that factor V Leiden carriers can clearly be delineated from normal individuals with the Coatest APCR test. A ratio of 2.0 as the cut-off provides 99.0% sensitivity and 95.4% specificity.To evaluate the lupus anticoagulant interference, we retrospectively analysed 16 lupus anticoagulant-positive patients. In this study, two (12.5%) showed a false-positive activated protein C resistance result. Six out of 16 (37.5%) lupus anticoagulant-positive patients were also carriers of the factor V Leiden mutation. Four out of eight (50%) false-positive activated protein C resistance results presented with an abnormal baseline clotting time. In order to prevent reporting false-positive results, a maximum baseline clotting time (65.8 s) was calculated. A new scheme for interpreting activated protein C resistance ratios was proposed.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验