Buckley S E, Ali P A, Evans C J, El-Sharkawi A M
School of Health Science, University of Wales Swansea, Singleton Park, Swansea, SA2 8PP, UK.
Phys Med Biol. 2006 Mar 7;51(5):1325-32. doi: 10.1088/0031-9155/51/5/020. Epub 2006 Feb 15.
Platinum enriched with 194Pt was irradiated for 4 days in NRG's TIRO 1 reactor, to produce (195m)Pt. Spectral analysis of the product was performed using a calibrated hyper pure germanium detector and its constituent radioisotopes were identified as (195m)Pt, 199Au and 192Ir. Using the detector's intrinsic efficiency calibration, their activities were estimated to be 1049, 133 and 5.8 MBq, respectively. The performance of the gamma camera was tested using quality control procedures recommended by the Institute of Physics and Engineering in Medicine (IPEM) and was found to be satisfactory. A torso phantom was used to determine the minimum detection limit (MDL) of (195m)Pt in a 2 cm diameter tumour using SPECT acquisitions (32 steps, 60 s per step). The MDL was found to be 8 ppm assuming an administered patient dose of 50 MBq and a total cisplatin dose of 105 mg. This work indicates that (195m)Pt-cisplatin is suitable for clinical scintigraphy and has led to the development of a clinical protocol that has been approved for a pilot study.
富含194Pt的铂在NRG的TIRO 1反应堆中辐照4天,以产生(195m)Pt。使用校准的超纯锗探测器对产物进行光谱分析,其组成放射性同位素被鉴定为(195m)Pt、199Au和192Ir。利用探测器的本征效率校准,估计它们的活度分别为1049、133和5.8 MBq。使用医学物理与工程学会(IPEM)推荐的质量控制程序对伽马相机的性能进行了测试,结果令人满意。使用躯干模型通过SPECT采集(32步,每步60秒)确定直径2 cm肿瘤中(195m)Pt的最小检测限(MDL)。假设给予患者的剂量为50 MBq,顺铂总剂量为105 mg,MDL为8 ppm。这项工作表明(195m)Pt-顺铂适用于临床闪烁扫描,并促成了一项已获批准用于初步研究的临床方案的制定。