钆喷酸葡胺(马根维显)、钆布醇(佳迪显)和钆贝葡胺(多它灵)在人血浆中于0.2、1.5和3特斯拉时的弛豫率
Relaxivity of Gadopentetate Dimeglumine (Magnevist), Gadobutrol (Gadovist), and Gadobenate Dimeglumine (MultiHance) in human blood plasma at 0.2, 1.5, and 3 Tesla.
作者信息
Pintaske Jörg, Martirosian Petros, Graf Hansjörg, Erb Gunter, Lodemann Klaus-Peter, Claussen Claus D, Schick Fritz
机构信息
Section on Experimental Radiology, Department of Diagnostic Radiology, University Hospital Tübingen, Tübingen, Germany.
出版信息
Invest Radiol. 2006 Mar;41(3):213-21. doi: 10.1097/01.rli.0000197668.44926.f7.
OBJECTIVES
We sought to determine the relaxivity and accurate relaxation rates of Gd-DTPA, Gd-BT-DO3A, and Gd-BOPTA at 0.2, 1.5, and 3 T in human blood plasma.
MATERIALS AND METHODS
Contrast media concentrations between 0.01 and 16 mM in human plasma were used for relaxation measurements. The R1 and R2 relaxation rates and r1 and r2 relaxivities were determined.
RESULTS
Gd-BOPTA produced the highest relaxation rates and relaxivities at all field strengths. The r1 and r2 values for Gd-BOPTA were 107-131% and 91-244% higher than for Gd-DTPA, respectively, and 72-98% and 82-166% higher than for Gd-BT-DO3A. Higher field strengths resulted in lower values of R1, R2, and r1 for all contrast agents tested and of r2 for Gd-DTPA and Gd-BT-DO3A. A linear dependence of R1 and R2 on concentration was found for Gd-DTPA and Gd-BT-DO3A and a nonlinear dependence for Gd-BOPTA for concentrations larger than 1 mM. The r1 and r2 relaxivity of Gd-BOPTA increased with decreasing concentration.
CONCLUSIONS
Gd-BOPTA demonstrates the highest longitudinal r1 at all field strengths, which is ascribable to weak protein interaction. The R2/R1 ratio increases at higher field strength only for Gd-BOPTA, hence very short echo times are required for Gd-BOPTA to benefit from the higher longitudinal relaxivity.
目的
我们试图测定钆喷酸葡胺(Gd - DTPA)、钆贝葡胺(Gd - BT - DO3A)和钆布醇(Gd - BOPTA)在0.2、1.5和3 T场强下于人体血浆中的弛豫率和准确的弛豫速率。
材料与方法
使用人体血浆中浓度介于0.01至16 mM之间的造影剂进行弛豫测量。测定了R1和R2弛豫速率以及r1和r2弛豫度。
结果
在所有场强下,钆布醇产生的弛豫速率和弛豫度最高。钆布醇的r1和r2值分别比钆喷酸葡胺高107 - 131%和91 - 244%,比钆贝葡胺高72 - 98%和82 - 166%。对于所有测试的造影剂,更高的场强导致R1、R2和r1值降低,对于钆喷酸葡胺和钆贝葡胺,r2值也降低。发现钆喷酸葡胺和钆贝葡胺的R1和R2与浓度呈线性关系,而对于浓度大于1 mM的钆布醇,R1和R2与浓度呈非线性关系。钆布醇的r1和r2弛豫度随浓度降低而增加。
结论
钆布醇在所有场强下均表现出最高的纵向r1,这归因于其与蛋白质的弱相互作用。仅钆布醇在更高场强下R2/R1比值增加,因此钆布醇需要非常短的回波时间才能从更高的纵向弛豫度中获益。