使用电子射野影像对两种头颈部固定装置进行的比较评估。

A comparative evaluation of two head and neck immobilization devices using electronic portal imaging.

作者信息

Donato K, Leszczynski K, Fleming K

机构信息

Northeastern Ontario Regional Cancer Centre, Hôpital régional de Sudbury Regional Hospital, 41 Ramsey Lake Road, Sudbury, Ontario, P3E 5J1, Canada.

出版信息

Br J Radiol. 2006 Feb;79(938):158-61. doi: 10.1259/bjr/32191494.

Abstract

A study was performed to compare the positioning reproducibility and the cost efficiency for two head and neck immobilization devices: the Uvex (Uvex Safety, Smithfield, USA) plastic mask system and the Finesse Frame with Ultraplast System (PLANET Medical, Svendborg, Denmark). 20 patients treated with 3D conformal radiation therapy for head and neck cancers were randomly selected (10 for each of the two different immobilization systems) and electronic portal images acquired during their course of treatment were saved and used in this study. The anatomical landmark coordinates and their shifts in the anteroposterior (AP) and craniocaudal (CC) directions with respect to the digitized simulator films for lateral fields were analysed using an in-house developed portal image registration system. Statistically, no evidence was found to indicate that the systematic components of the displacement for the Uvex system and the Finesse Frame with Ultraplast System were different from each other or from zero. The random component of displacement was slightly smaller in the AP direction for the Uvex than the Ultraplast system (sigma = 1.9 mm and 2.9 mm, respectively, p = 0.007), but larger in the CC direction (sigma = 3.8 mm and 2.2 mm, respectively, p<10(-9)). Production time and required materials for a radiation therapy department were also quantified to assess costs for each system. The overall costs per patient were estimated at 141.50 dollars (CAD) and 82.10 dollars for the Uvex and Ultraplast systems, respectively. The Finesse Frame with Ultraplast System of immobilization for head and neck cancer treatment provides a field placement reproducibility that is equal to, or greater than, that of the Uvex plastic mask immobilization system and, while it requires more expensive materials, the workload and consequently overall cost is greatly reduced.

摘要

开展了一项研究,以比较两种头颈部固定装置的定位重复性和成本效益:优唯斯(Uvex Safety,美国史密斯菲尔德)塑料面罩系统和配备超塑系统的Finesse框架(丹麦斯文堡PLANET Medical公司)。随机选取20例接受头颈部癌症三维适形放射治疗的患者(两种不同固定系统各10例),保存其治疗过程中采集的电子门静脉图像并用于本研究。使用自行开发的门静脉图像配准系统,分析了相对于侧野数字化模拟片在前后(AP)和颅尾(CC)方向上的解剖标志坐标及其移位情况。从统计学角度来看,没有证据表明优唯斯系统和配备超塑系统的Finesse框架的位移系统成分彼此不同或与零有差异。优唯斯系统在AP方向上的随机位移成分略小于超塑系统(标准差分别为1.9毫米和2.9毫米,p = 0.007),但在CC方向上更大(标准差分别为3.8毫米和2.2毫米,p<10⁻⁹)。还对放射治疗科室的生产时间和所需材料进行了量化,以评估每个系统的成本。优唯斯系统和超塑系统的每位患者总成本估计分别为141.50加元(CAD)和82.10加元。用于头颈部癌症治疗的配备超塑系统的Finesse框架在射野放置重复性方面与优唯斯塑料面罩固定系统相当或更高,虽然其所需材料更昂贵,但工作量和总成本大幅降低。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验