Valaskovic Gary A, Utley Lucas, Lee Mike S, Wu Jing-Tao
New Objective, Inc., Woburn, MA 01801, USA.
Rapid Commun Mass Spectrom. 2006;20(7):1087-96. doi: 10.1002/rcm.2414.
Nanospray experiments were performed on an ensemble of drug molecules and their commonly known metabolites to compare performance with conventional electrospray ionization (ESI) and to evaluate equimolar response capabilities. Codeine, dextromethorphan, tolbutamide, phenobarbital, cocaine, and morphine were analyzed along with their well-known metabolites that were formed via hydroxylation, dealkylation, hydrolysis, and glucuronidation. Nanospray exhibited a distinct trend toward equimolar response when flow rate was reduced from 25 nL/min to less than 10 nL/min. A more uniform response between the parent drug and the corresponding metabolites was obtained at flow rates of 10 nL/min or lower. The largest discrepancy was within +/-50% for plasma samples. Nanospray was used as a calibrator for conventional ESI liquid chromatography/tandem mass spectrometry (LC/MS/MS) and normalization factors were applied to the quantitation of an acyl-glucuronide metabolite of a proprietary compound in rat plasma. A nanospray calibration method was developed with the standard curve of the parent drug to generate quantitative results for drug metabolites within +/-20% of that obtained with reference standards and conventional ESI. The nanospray method provides a practical solution for the quantitative estimation of drug metabolites in drug discovery when reference standards are not available.
对一组药物分子及其常见代谢物进行了纳米喷雾实验,以与传统电喷雾电离(ESI)比较性能,并评估等摩尔响应能力。对可待因、右美沙芬、甲苯磺丁脲、苯巴比妥、可卡因和吗啡及其通过羟基化、脱烷基化、水解和葡萄糖醛酸化形成的知名代谢物进行了分析。当流速从25纳升/分钟降低到小于10纳升/分钟时,纳米喷雾呈现出明显的等摩尔响应趋势。在10纳升/分钟或更低的流速下,母体药物和相应代谢物之间获得了更均匀的响应。血浆样品的最大差异在±50%以内。纳米喷雾被用作传统ESI液相色谱/串联质谱(LC/MS/MS)的校准剂,并将归一化因子应用于大鼠血浆中一种专利化合物的酰基葡萄糖醛酸代谢物的定量分析。利用母体药物的标准曲线开发了一种纳米喷雾校准方法,以生成药物代谢物的定量结果,其与参考标准品和传统ESI获得的结果相差±20%以内。当没有参考标准品时,纳米喷雾方法为药物发现中药物代谢物的定量估计提供了一种实用的解决方案。