Shively M L
University of Colorado School of Pharmacy, Boulder 80309.
J Pharm Sci. 1991 Apr;80(4):376-9. doi: 10.1002/jps.2600800419.
The use of mercury porosimetry to determine pore size and pore distribution is currently applied in a variety of areas where powders or membranes are utilized. Conventional analysis of mercury porosimetry data assumes a cylindrical pore configuration. The method of analysis presented in this study requires no geometrical assumptions to be made and provides a means to graphically delineate between different pore geometries. This method of analysis also provides data that shows evidence for the existence of different pore shapes within the same sample, where only cylindrical pores have been assumed in the past. Furthermore, this method of analysis makes it possible to evaluate intrusion and extrusion data and to postulate as to the validity of contact angle and volume hysteresis.
压汞法用于测定孔径和孔隙分布,目前在各种使用粉末或膜的领域中都有应用。传统的压汞法数据分析假定孔隙结构为圆柱形。本研究提出的分析方法无需进行几何假设,并提供了一种以图形方式区分不同孔隙几何形状的手段。这种分析方法还提供了数据,表明在同一样品中存在不同的孔隙形状,而过去仅假定为圆柱形孔隙。此外,这种分析方法使得评估侵入和挤出数据以及推断接触角和体积滞后的有效性成为可能。