Wan Shuying, Joshi Chandra P, Carnes Greg, Schreiner L John
Cancer Centre of Southeastern Ontario, 25 King Street West, Kingston, Ontario K7L 5P9, Canada.
J Appl Clin Med Phys. 2006 Winter;7(1):115-25. doi: 10.1120/jacmp.v7i1.2114. Epub 2006 Feb 15.
Automated seed loaders for permanent prostate implants are now commercially available. Besides improved radiation safety, these systems offer seed assay capability and ease of needle loading, making preplanned as well as intra-operative implant procedures more time-efficient. The Isoloader (Mentor Corp., CA) uses individual I125 seeds (SL-125 ProstaSeed) loaded in up to 199 chambers inside a shielded cartridge. The unit performs seed counting and calibration using a builtin solid-state detector. In order to evaluate the reproducibility and accuracy of the calibration process, two test cartridges were measured with the Isoloader itself and compared with a well-type ionization chamber (HDR-1000Plus, Standard Imaging). The air kerma strength measurements for all seeds using the Isoloader had a standard deviation of about 2.7%. For the eight seeds assayed more intensively using both the Isoloader and well chamber, the standard deviations of the measurements for each seed were in the range of 0.8% to 2.8% and 0.6% to 1.3%, respectively. The variation in the Isoloader calibration is attributed to small detector solid angle and bead geometry within seed capsules (verified by radiographs). The reproducibility of the air kerma strength measured by the Isoloader was comparable to that from the well chamber and was clinically acceptable. Seed strength measured with the Isoloader was on average 1% 2% larger than that measured with the well chamber, indicating that the accuracy of the Isoloader was clinically acceptable.
用于永久性前列腺植入的自动种子源加载器现已商业化。除了提高辐射安全性外,这些系统还具备种子源分析能力且易于装载针,使术前以及术中植入程序更高效。Isoloader(美国加利福尼亚州的Mentor公司)使用单个I125种子(SL - 125 ProstaSeed),这些种子被装载在屏蔽盒内多达199个腔室中。该装置使用内置的固态探测器进行种子计数和校准。为了评估校准过程的可重复性和准确性,使用Isoloader本身对两个测试盒进行了测量,并与井型电离室(HDR - 1000Plus,标准成像公司)进行比较。使用Isoloader对所有种子进行的空气比释动能强度测量的标准偏差约为2.7%。对于同时使用Isoloader和井型电离室进行更密集分析的8颗种子,每颗种子测量的标准偏差分别在0.8%至2.8%和0.6%至1.3%的范围内。Isoloader校准的变化归因于探测器的小立体角和种子胶囊内的珠子几何形状(经射线照片验证)。Isoloader测量的空气比释动能强度的可重复性与井型电离室相当,在临床上是可接受的。用Isoloader测量的种子强度平均比用井型电离室测量的大1%至2%,这表明Isoloader的准确性在临床上是可接受的。