Li Zuofeng
Department of Radiation Oncology, Mallinckrodt Institute of Radiology, Washington University School of Medicine, St Louis, Missouri 63110, USA.
Med Phys. 2002 Jun;29(6):1029-34. doi: 10.1118/1.1478559.
This report presents the results of Monte Carlo calculations of the dosimetric parameters of the Urocor ProstaSeed 1251 seed source. This source contains five spherical silver markers, of 0.5 mm diameter, with 1251 deposited on the spheres through ion exchange. The silver spheres are then encapsulated within a 0.8 mm in diameter and 4.5 mm long cylindrical shell of titanium, as is common to this type of sources. Physical dimensions of the source are confirmed by measurement. Four (4) geometric models of the source, based on different assumptions on the locations of the silver spheres within the seed, were used in dose rate calculations. Monte Carlo photon transport simulation was used to calculate the dosimetric parameters of dose rate constant, radial dose function, and anisotropy function using these geometric models of the source. The Monte Carlo calculated dose rate constant of the ProstaSeed source was found equal to 0.925 cGy/Uh with approximate uncertainties of 5%. Radial dose function values and anisotropy function values were derived from the relative dose distribution around the source calculated by the Monte Carlo code. The calculated values of these dosimetric parameters agree with previously published thermoluminescence-dosimeter-measured values for this source after consideration of measurement and calculation uncertainties.
本报告展示了对Urocor ProstaSeed 125I种子源剂量学参数进行蒙特卡罗计算的结果。该源包含五个直径为0.5毫米的球形银标记物,通过离子交换将125I沉积在这些球体上。然后,银球被封装在一个直径0.8毫米、长4.5毫米的钛制圆柱形外壳内,这是此类源的常见做法。通过测量确认了源的物理尺寸。在剂量率计算中使用了基于对种子内银球位置的不同假设的四种(4种)源几何模型。利用这些源几何模型,通过蒙特卡罗光子输运模拟计算剂量率常数、径向剂量函数和各向异性函数的剂量学参数。发现ProstaSeed源的蒙特卡罗计算剂量率常数等于0.925 cGy/(U·h),近似不确定度为5%。径向剂量函数值和各向异性函数值源自蒙特卡罗代码计算的源周围相对剂量分布。在考虑测量和计算不确定度后,这些剂量学参数的计算值与该源先前发表的热释光剂量计测量值一致。