Burel-Deschamps Laure, Giamarchi Philippe, Stephan Ludovic, Lijour Yves, Le Bihan Alain
UMR CNRS 6521, Université de Bretagne Occidentale, 6 avenue Le Gorgeu, BP 809, 29285, Brest Cedex, France.
J Fluoresc. 2006 Mar;16(2):177-83. doi: 10.1007/s10895-005-0044-x. Epub 2006 Mar 17.
The absorption and fluorescence spectra of carbaryl (CB), carbofuran (CF) and carbendazim (MBC) have been studied. Fluorescence lifetime and fluorescence quantum yields are also reported as well as the influence of pH, solvent and presence of humic acids on fluorescence. The limit of detection (LD) of the three compounds has been measured by direct analysis by laser-induced fluorescence (LIF) using a pulsed YAG laser with an Optical Parametric Oscillator (OPO) as excitation source and an Intensified Charged Coupled Device (ICCD) camera for the fluorescence detection. Instrumental LD found for CB, for MBC and for CF are respectively 4, 50 and 1000 ng L(-1). In tap water, the LD obtained is 800 ng L(-1) for MBC and 20,000 ng L(-1) for CF. For CB, the use of a time shift between excitation and emission allows to reach a LD of 20 ng L(-1) in tap water.
研究了西维因(CB)、克百威(CF)和多菌灵(MBC)的吸收光谱和荧光光谱。还报告了荧光寿命和荧光量子产率,以及pH值、溶剂和腐殖酸的存在对荧光的影响。使用带有光学参量振荡器(OPO)的脉冲YAG激光器作为激发源,以及用于荧光检测的增强型电荷耦合器件(ICCD)相机,通过激光诱导荧光(LIF)直接分析测量了这三种化合物的检测限(LD)。CB、MBC和CF的仪器检测限分别为4、50和1000 ng L(-1)。在自来水中,MBC的检测限为800 ng L(-1),CF的检测限为20,000 ng L(-1)。对于CB,通过在激发和发射之间使用时间偏移,可在自来水中达到20 ng L(-1)的检测限。