可编程脑脊液分流阀:3T 场强下的磁共振成像安全性体外评估

Programmable CSF shunt valve: in vitro assessment of MR imaging safety at 3T.

作者信息

Shellock F G, Habibi R, Knebel J

机构信息

Keck School of Medicine, University of Southern California and Institute for Magnetic Resonance Safety, Education, and Research, Los Angeles, Calif 90045, USA.

出版信息

AJNR Am J Neuroradiol. 2006 Mar;27(3):661-5.

Abstract

A programmable CSF shunt valve was assessed for magnetic field interactions, heating (transmit-receive body radio-frequency coil; whole-body averaged specific absorption rate, 2.1 W/kg), functional alterations, and artifacts at 3T. The programmable valve showed minor magnetic field interactions and heating was not excessive (+0.8 degrees C). The function of the programmable valve was not altered by multiple exposures to the 3T scanner or from exposure to various MR imaging conditions. Therefore, this implant is safe for a patient undergoing MR imaging at 3T or less when the radiologist follows specific safety guidelines. Artifacts for the programmable valve were relatively large in relation to the size and shape of the valve; this finding may impact the diagnostic use of MR imaging if the area of interest is in proximity to this implant.

摘要

对一种可编程脑脊液分流阀进行了评估,以确定其在3T磁场下的磁场相互作用、发热情况(发射-接收体射频线圈;全身平均比吸收率,2.1W/kg)、功能改变及伪影。该可编程阀显示出轻微的磁场相互作用,发热也不过高(+0.8摄氏度)。多次暴露于3T扫描仪或不同的磁共振成像条件下,可编程阀的功能未发生改变。因此,当放射科医生遵循特定的安全指南时,对于接受3T及以下磁共振成像检查的患者,这种植入物是安全的。该可编程阀的伪影相对于阀的大小和形状来说比较大;如果感兴趣区域靠近该植入物,这一发现可能会影响磁共振成像的诊断应用。

相似文献

2
Magnetically programmable shunt valve: MRI at 3-Tesla.
Magn Reson Imaging. 2007 Sep;25(7):1116-21. doi: 10.1016/j.mri.2006.12.004. Epub 2007 Jan 19.
3
Assessment of MRI issues for a 3-T "immune" programmable CSF shunt valve.
AJR Am J Roentgenol. 2011 Jul;197(1):202-7. doi: 10.2214/AJR.10.5915.
4
Adjustable cerebrospinal fluid shunt valves in 3.0-Tesla MRI: a phantom study using explanted devices.
Rofo. 2010 Jul;182(7):594-602. doi: 10.1055/s-0028-1109882. Epub 2009 Dec 2.
7
Assessment of MRI issues for a new cerebral spinal fluid shunt, gravitational valve (GV).
Magn Reson Imaging. 2017 Dec;44:8-14. doi: 10.1016/j.mri.2017.07.018. Epub 2017 Jul 20.
8
Programmable shunts and headphones: Are they safe together?
J Neurosurg Pediatr. 2015 Oct;16(4):402-5. doi: 10.3171/2015.1.PEDS14400. Epub 2015 Jul 7.
9
Effect of 3T MRI on the function of shunt valves--evaluation of Paedi GAV, Dual Switch and proGAV.
Eur J Radiol. 2005 Oct;56(1):56-9. doi: 10.1016/j.ejrad.2005.03.029.

引用本文的文献

1
Measuring CSF shunt flow with MRI using flow enhancement of signal intensity (FENSI).
Magn Reson Med. 2024 Aug;92(2):807-819. doi: 10.1002/mrm.30079. Epub 2024 Mar 12.
2
Development of shunt valves used for treating hydrocephalus: comparison with endoscopy treatment.
Childs Nerv Syst. 2023 Oct;39(10):2709-2717. doi: 10.1007/s00381-023-06049-1. Epub 2023 Jun 24.
5
Effect of orthodontic brackets and different wires on radiofrequency heating and magnetic field interactions during 3-T MRI.
Dentomaxillofac Radiol. 2014;43(2):20130356. doi: 10.1259/dmfr.20130356. Epub 2013 Nov 20.
6
Evaluation of magnetic resonance imaging issues for implantable microfabricated magnetic actuators.
Biomed Microdevices. 2014 Feb;16(1):153-61. doi: 10.1007/s10544-013-9815-3.
7
3T magnetic resonance imaging testing of externally programmable shunt valves.
Surg Neurol Int. 2012;3:81. doi: 10.4103/2152-7806.99171. Epub 2012 Jul 28.
8
Programmable CSF shunt valves: radiographic identification and interpretation.
AJNR Am J Neuroradiol. 2010 Aug;31(7):1343-6. doi: 10.3174/ajnr.A1997. Epub 2010 Feb 11.
9
Ventricular assist device implant (AB 5000) prototype cannula: in vitro assessment of MRI issues at 3-Tesla.
J Cardiovasc Magn Reson. 2008 May 21;10(1):23. doi: 10.1186/1532-429X-10-23.

本文引用的文献

1
Adjustment and malfunction of a programmable valve after exposure to toy magnets. Case report.
J Neurosurg. 2004 Nov;101(2 Suppl):222-5. doi: 10.3171/ped.2004.101.2.0222.
3
Reliability of a new adjustable shunt device without the need for readjustment following 3-Tesla MRI.
Childs Nerv Syst. 2005 Mar;21(3):227-9. doi: 10.1007/s00381-004-1084-2. Epub 2005 Jan 19.
4
Implantable microstimulator: magnetic resonance safety at 1.5 Tesla.
Invest Radiol. 2004 Oct;39(10):591-9. doi: 10.1097/01.rli.0000138090.43450.ec.
5
MR procedures: biologic effects, safety, and patient care.
Radiology. 2004 Sep;232(3):635-52. doi: 10.1148/radiol.2323030830. Epub 2004 Jul 29.
6
Biomedical implants and devices: assessment of magnetic field interactions with a 3.0-Tesla MR system.
J Magn Reson Imaging. 2002 Dec;16(6):721-32. doi: 10.1002/jmri.10207.
7
Electromagnetic field hazards involving adjustable shunt valves in hydrocephalus.
J Neurosurg. 2002 Feb;96(2):331-4. doi: 10.3171/jns.2002.96.2.0331.
9
Hydrodynamic properties of hydrocephalus shunts.
Acta Neurochir Suppl. 1998;71:334-9. doi: 10.1007/978-3-7091-6475-4_97.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验