During pacemaker implantation in a patient with permanent atrial fibrillation, it remained impossible to advance a passive fixation lead with fins through the right atrium. However, a lead with a retractable screw easily passed the right atrium and was positioned in the right ventricle. Transesophageal echocardiography revealed an extensive net-like perforated Eustachian valve within the right atrium that had caused entrapment of the anchor fins during lead implantation. Remnants of embryonal structures within the right atrium should be considered a rare possible barrier during pacemaker implantation.