Indirect calorimetry: methodology, instruments and clinical application.

作者信息

da Rocha Eduardo E Moreira, Alves Valéria Girard F, da Fonseca Rosana Barcellos V

机构信息

NUTROCLIN, Clínica São Vicente/Gávea, Rio de Janeiro, Brazil.

出版信息

Curr Opin Clin Nutr Metab Care. 2006 May;9(3):247-56. doi: 10.1097/01.mco.0000222107.15548.f5.

Abstract

PURPOSE OF REVIEW

This review aims to identify the basic methods for accurately measuring a patient's energy expenditure in clinical nutrition practice by indirect calorimetry, and the impact upon a disease state of applying the results obtained.

RECENT FINDINGS

The open-circuit method is the most widely used in the majority of classical instruments for measuring energy consumption. Advances in gas exchange measurement have made this technique readily and precisely available at the bedside. Nevertheless, it is important to understand its intricate primary methodology for safe and correct application. The stress and activity factors should be carefully and specifically applied, and the respiratory quotient abandoned, for tailoring a patient's daily nutrition regimens. Caloric expenditure measured by indirect calorimetry coupled with the doubly labeled water technique introduced the concept of physical activity energy expenditure, which added to resting energy expenditure results in total daily energy expenditure. Compact modular and handheld devices have been introduced into the market, together with similar technology for evaluating exercise energy expenditure, making utilization easier, safer and precise. In the critically ill population, which is exposed to medical and surgical interventions, indirect calorimetry has greatly changed the practice of caloric administration, significantly reducing the total daily amount.

SUMMARY

In conclusion, one has to be careful when choosing devices, and understanding and clinically applying the results obtained by indirect calorimetry, bearing in mind that measured resting energy expenditure should be the daily caloric goal in order to diminish clinical morbidity.

摘要

综述目的

本综述旨在确定通过间接测热法在临床营养实践中准确测量患者能量消耗的基本方法,以及应用所得结果对疾病状态的影响。

最新发现

开路法在大多数经典能量消耗测量仪器中应用最为广泛。气体交换测量技术的进步使该技术能够在床边轻松、精确地实现。然而,为了安全、正确地应用,了解其复杂的基本方法很重要。应仔细、具体地应用应激和活动因素,并摒弃呼吸商,以制定患者的每日营养方案。间接测热法测量的热量消耗与双标记水技术相结合,引入了身体活动能量消耗的概念,将其与静息能量消耗相加得出每日总能量消耗。紧凑型模块化和手持式设备已投放市场,同时还有评估运动能量消耗的类似技术,使应用更加简便、安全和精确。在接受医疗和外科干预的危重症人群中,间接测热法极大地改变了热量管理的实践,显著减少了每日总量。

总结

总之,选择设备时必须谨慎,理解并临床应用间接测热法所得结果,同时要牢记,为降低临床发病率,测量的静息能量消耗应作为每日热量目标。