Schwartz H E, Ulfelder K, Sunzeri F J, Busch M P, Brownlee R G
Beckman Instruments, Inc., Palo Alto, CA 94304.
J Chromatogr. 1991 Oct 18;559(1-2):267-83. doi: 10.1016/0021-9673(91)80077-t.
A high-performance capillary electrophoresis system with a polysiloxane-coated capillary and polymeric buffer additives was investigated for the analysis of DNA restriction fragments and polymerase chain reaction (PCR) products. Mobility data and Ferguson plots of the DNA fragments at different polymer (hydroxypropylmethylcellulose) concentrations indicated that effective molecular sieving was obtained consistent with existing data of conventional gel electrophoresis and with recent HPCE data. The precision and peak efficiency were excellent and the system was applied to the analysis of specific co-amplified DNA sequences (HIV-1 and HLA-DQ-alpha). After PCR, ultrafiltration was used in the sample preparation step to desalt the sample and to remove superfluous PCR reaction products. Electrokinetic injection was used for sample introduction into the capillary. The addition of ethidium bromide to the buffer resulted in longer migration times of DNA fragments and better peak resolution. During HPCE, an artifact associated with dilute DNA solutions leading to the appearance of extra peaks in the electropherogram was found.
研究了一种配备聚硅氧烷涂层毛细管和聚合物缓冲添加剂的高效毛细管电泳系统,用于分析DNA限制性片段和聚合酶链反应(PCR)产物。不同聚合物(羟丙基甲基纤维素)浓度下DNA片段的迁移率数据和弗格森图表明,有效分子筛分与传统凝胶电泳的现有数据以及最近的高效毛细管电泳数据一致。精密度和峰效率极佳,该系统用于分析特定的共扩增DNA序列(HIV-1和HLA-DQ-α)。PCR后,在样品制备步骤中使用超滤对样品进行脱盐并去除多余的PCR反应产物。采用电动进样将样品引入毛细管。向缓冲液中添加溴化乙锭导致DNA片段的迁移时间延长,峰分辨率提高。在高效毛细管电泳过程中,发现了一种与稀DNA溶液相关的假象,导致电泳图中出现额外的峰。