Swennen Gwen R J, Schutyser Filip, Barth Enno-Ludwig, De Groeve Pieter, De Mey Albert
Department of Plastic Surgery, University Hospital Brugmann, ULB, Brussels, Belgium.
J Craniofac Surg. 2006 Mar;17(2):314-25. doi: 10.1097/00001665-200603000-00019.
The purpose of this study was to present a new innovative three-dimensional (3-D) cephalometric method. Part I deals with the set-up and validation of a voxel-based semi-automatic 3-D cephalometric reference system. The CT data (DICOM 3.0 files) of 20 control patients with normal skeletal relationships were used for this study. To investigate accuracy and reliability of the 3-D cephalometric reference system (Maxilimtrade mark, version 1.3.0) a total of 42 (14 horizontal, 14 vertical and 14 transversal) orthogonal measurements were performed on each patient twice by each of two investigators. The intra-observer measurement error was less then 0.88 mm, 0.76 mm and 0.84 mm for horizontal, vertical and transversal orthogonal measurements, respectively. The inter-observer measurement error was less as 0.78 mm, 0.86 mm and 1.26 mm for horizontal, vertical and transversal orthogonal measurements, respectively. Squared correlation coefficients showed a high intra-observer and inter-observer reliability. The presented 3-D cephalometric reference system proved to be accurate and reliable and can therefore be used for 3-D cephalometric hard and soft tissue analysis.
本研究的目的是提出一种全新的三维(3-D)头影测量方法。第一部分涉及基于体素的半自动三维头影测量参考系统的建立与验证。本研究使用了20例骨骼关系正常的对照患者的CT数据(DICOM 3.0文件)。为了研究三维头影测量参考系统(Maxilim商标,版本1.3.0)的准确性和可靠性,两名研究者分别对每位患者进行了两次共42项(14项水平、14项垂直和14项横向)正交测量。观察者内测量误差在水平、垂直和横向正交测量中分别小于0.88毫米、0.76毫米和0.84毫米。观察者间测量误差在水平、垂直和横向正交测量中分别小于0.78毫米、0.86毫米和1.26毫米。平方相关系数显示出较高的观察者内和观察者间可靠性。所提出的三维头影测量参考系统被证明是准确可靠的,因此可用于三维头影测量的硬组织和软组织分析。