Suppr超能文献

Coronary sinus stenting for the stabilization of left ventricular lead during resynchronization therapy.

作者信息

Kowalski Oskar, Prokopczuk Janusz, Lenarczyk Radoslaw, Pruszkowska-Skrzep Patrycja, Polonski Lech, Kalarus Zbigniew

机构信息

First Department of Cardiology, Silesian Medical School, Silesian Center for Heart Diseases, Zabrze, Szpitalna 2, Poland.

出版信息

Europace. 2006 May;8(5):367-70. doi: 10.1093/europace/eul022.

Abstract

We report on two patients treated with cardiac resynchronization therapy, in whom early (intra-operatively, 64-year-old man) and late (4 months post-operatively, 57-year-old woman) instability of the left ventricular (LV) lead occurred. In order to stabilize the electrodes, stents were deployed in both patients within the coronary sinus, into the space between the lead and the wall of the vein effectively pinning the lead to the wall. During 3 and 5 months of follow-up, the electrodes remained stable and allowed for successful resynchronization in both cases. Stenting within the coronary sinus seems to be a safe method for LV lead stabilization, which can substantially increase the success rate of resynchronization therapy. This new approach, although promising, has to prove its safety and should not be practised routinely until long-term follow-up data are available.