Onori S, Bortolin E, Calicchia A, Carosi A, De Angelis C, Grande S
Department of Technology and Health, Istituto Superiore di Sanità, Viale Regina Elena, 299, 00161 Rome, Italy.
Radiat Prot Dosimetry. 2006;120(1-4):226-9. doi: 10.1093/rpd/nci632. Epub 2006 Apr 27.
In the implementation of a large-scale dosimetry intercomparison one of the main constraints is the availability of large number of dosemeters of the highest quality. Therefore, ISS tested the possibility of using commercially available dosemeters, alanine pellets and thermoluminescence (TL) dosimetry chips, for transfer dosimetry within the Italian intercomparison programme. In this work the characterisation of commercial alanine and TL dosemeters along with the ISS dose assessment procedure used in the Italian intercomparison are reported. Results demonstrate the feasibility of the ISS approach to transfer dosimetry since it is possible to measure 10 Gy with a combined uncertainty of 1% (1sigma) and 1 Gy with a combined uncertainty of 1.7% (1sigma) with alanine and TL dosemeters, respectively.