O'Gorman Ruth L, Summers Paul E, Zelaya Fernando O, Williams Steven C R, Alsop David C, Lythgoe David J
King's College London, Institute of Psychiatry, Centre for Neuroimaging Sciences, UK.
Magn Reson Med. 2006 Jun;55(6):1291-7. doi: 10.1002/mrm.20864.
The accurate quantification of perfusion with arterial spin labeling (ASL) requires consideration of a number of factors, including the efficiency of the inversion and control pulses used for spin labeling. In this study the effects of spin velocity on continuous ASL efficiency when using the amplitude modulated control strategy were investigated using simulations of the Bloch equations. The inversion efficiency was determined in vivo by combining the simulations with phase-contrast velocity mapping data acquired at the level of the tagging plane. Using this novel method, an average inversion efficiency of 69% was calculated for a group of 28 subjects, in good agreement with experimental data reported previously. There was, however, a large range in inversion efficiency measured across the subject group (50-76%), indicating that the velocity dependence of the amplitude modulated control efficiency may introduce additional variability into the perfusion calculations if not properly taken into account.
使用动脉自旋标记(ASL)进行灌注的准确量化需要考虑许多因素,包括用于自旋标记的反转脉冲和控制脉冲的效率。在本研究中,使用布洛赫方程模拟研究了采用幅度调制控制策略时自旋速度对连续ASL效率的影响。通过将模拟与在标记平面水平获取的相位对比速度映射数据相结合,在体内确定反转效率。使用这种新方法,计算出一组28名受试者的平均反转效率为69%,与先前报道的实验数据高度一致。然而,在整个受试者组中测量到的反转效率范围很大(50 - 76%),这表明如果没有适当考虑,幅度调制控制效率对速度的依赖性可能会在灌注计算中引入额外的变异性。