A review of leakage flow in centrifugal blood pumps.

作者信息

Chan Weng-Kong, Wong Yew-Wah

机构信息

School of Mechanical & Aerospace Engineering, Nanyang Technological University, Singapore.

出版信息

Artif Organs. 2006 May;30(5):354-9. doi: 10.1111/j.1525-1594.2006.00225.x.

Abstract

This article presents a new approach in determining the functional relationship between the leakage flow in a centrifugal blood pump and various parameters that affect it. While high leakage flow in a blood pump is essential for good washout and can help prevent thrombus formation, excessive leakage flow will result in higher fluid shear stress that may lead to hemolysis. Dimensional analysis is employed to provide a functional relationship between leakage flow rate and other important parameters governing the operation of a centrifugal blood pump. Results showed that pump performance with a smaller gap clearance is clearly superior compared to those of two other similar pumps with larger gap clearances. It was also observed that the nondimensional leakage flow rate varies almost linearly with dimensionless pump head. It also decreases with increasing volume flow rate. A smaller gap clearance will also increase the flow resistance and hence, decrease the nondimensional leakage flow rate. Increasing surface roughness, length of the gap clearance passage, or loss coefficient of the gap geometry will increase losses and hence, decrease the leakage flow rate. It is also interesting to note that for a given pump and gap clearance geometry, the nondimensional leakage flow rate is almost independent of the Reynolds number when specific speed is constant.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验