Roberts M A, Tang C C
Department of Physics and Astronomy, University College London, Gower Street, London WC1E 6BT, UK.
J Synchrotron Radiat. 1998 Sep 1;5(Pt 5):1270-4. doi: 10.1107/S0909049598007870.
In an attempt to improve the resolution in flat-plate diffraction geometry, a new set of parallel foils, 100 microm spacing, 365 mm long with an aperture of 20 x 20 mm(2), has been commissioned. The study was carried out using the two-circle diffractometer of station 2.3 at the SRS. In order to properly quantify the improvements, a detailed and comparative study of the instrumental resolution using the new and existing foils, of 200 micron spacing, is reported. A number of cubic materials, BaF2, Si and CeO2, that are known to show well defined Bragg peaks over the full 2theta angular range available, were investigated.
为了提高平板衍射几何结构中的分辨率,已启用一组新的平行箔片,其间距为100微米,长度为365毫米,孔径为20×20毫米²。该研究使用了SRS的2.3号站的双圆衍射仪进行。为了正确量化改进效果,报告了一项使用新的和现有的间距为200微米的箔片对仪器分辨率进行的详细比较研究。研究了一些立方材料,如BaF₂、Si和CeO₂,已知这些材料在整个可用的2θ角范围内都能显示出清晰的布拉格峰。