Determination of phenylalanine in human serum by isotope dilution liquid chromatography/tandem mass spectrometry.

作者信息

Lee Hwashim, Park Sangryoul, Lee Gaeho

机构信息

Health Metrology Group, Division of Metrology for Quality Life, Korea Research Institute of Standards and Science, P.O. Box 102, Yuseong, Daejeon, 305-600, Republic of Korea.

出版信息

Rapid Commun Mass Spectrom. 2006;20(12):1913-7. doi: 10.1002/rcm.2528.

Abstract

Isotope dilution liquid chromatography/tandem mass spectrometry (ID-LC/MS/MS) has been developed as a candidate reference method to determine the level of phenylalanine in human serum. The advantages of this method include a simple sample preparation without derivatization, selective detection of analytes, and the use of an isotopic analogue as an internal standard. Phenylalanine and its isotopic analogue, phenylalanine-ring-(13)C(6), were monitored at the transitions m/z 166.2/120.2 and 172.2/126.2 in the multiple-reaction monitoring (MRM) mode, respectively. The expanded uncertainty of the measurement result of phenylalanine in the serum was approximately 1.2% within a 95% confidence level. A standard reference material, with a certified value of phenylalanine, was analyzed in order to verify this method. The result obtained by the ID-LC/MS/MS method differed somewhat from the certified value, but agreed well with the gravimetric value. The measurement result of phenylalanine in serum by ID-LC/MS/MS was compared with the results from the commercial HPLC method, which was carried out in clinics. The results from the commercial HPLC method showed inconsistent results with each other. The busted results from the commercial HPLC method suggest that it should be possible to trace the results of the commercial fields to well-characterized reference materials or methods.

摘要

同位素稀释液相色谱/串联质谱法(ID-LC/MS/MS)已被开发为一种用于测定人血清中苯丙氨酸水平的候选参考方法。该方法的优点包括无需衍生化的简单样品制备、对分析物的选择性检测以及使用同位素类似物作为内标。在多反应监测(MRM)模式下,分别在m/z 166.2/120.2和172.2/126.2的跃迁处监测苯丙氨酸及其同位素类似物苯丙氨酸-环-(13)C(6)。在95%置信水平下,血清中苯丙氨酸测量结果的扩展不确定度约为1.2%。为了验证该方法,分析了一种具有苯丙氨酸认证值的标准参考物质。ID-LC/MS/MS方法获得的结果与认证值略有不同,但与重量法值吻合良好。将ID-LC/MS/MS法测定血清中苯丙氨酸的结果与临床使用的商业HPLC法的结果进行了比较。商业HPLC法的结果相互不一致。商业HPLC法结果的不一致表明,将商业领域的结果追溯到特征明确的参考物质或方法应该是可行的。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验