钙片对腰椎双能X线吸收测定扫描结果判读的影响。

Effect of calcium tablets on interpretation of lumbar spine DXA scans.

作者信息

Kendler David L, Kiebzak Gary M, Ambrose Catherine G, Dinu C, Robertson Stephen, Schmeer Peg, Van Pelt Janet L

机构信息

Osteoporosis Research Centre, Vancouver, British Columbia, Canada.

出版信息

J Clin Densitom. 2006 Jan-Mar;9(1):97-104. doi: 10.1016/j.jocd.2005.12.002. Epub 2006 Mar 29.

Abstract

Undissolved calcium (Ca) tablets in the gastrointestinal tract at the time of a dual-energy X-ray absorptiometry (DXA) scan could conceivably affect the accuracy of the lumbar spine bone mineral density (BMD) determination. We studied phantoms and volunteers to determine the effect of Ca tablets overlying bone, Ca tablets in the soft-tissue field, and Ca tablets overlapping both bone and soft tissue. For L1-4, a 500-mg Ca tablet taped to the phantom surface or the skin of volunteers resulted in <or=2.2% mean increase in measured BMD (less than the lumbar spine least significant change), changing the T-score for L1-4 very little. However, an overlying Ca tablet had a substantial effect on BMD of a single vertebral body. Greater effects were seen with higher amounts of tablet Ca and with lower BMD, resulting in as much as a 12.6% mean increase in single vertebral BMD in the in vivo series. Recent versions of DXA software successfully identified tablets in the soft-tissue field as artifact and eliminated the tablet pixels from the analysis. However, tablets overlapping bone and soft tissue necessitated operator intervention to either exclude the affected vertebral body or adjust edges by mapping the tablet as artifact and/or neutral field. A paraffin-coated Ca tablet swallowed by a volunteer was tracked serially by DXA during gastrointestinal transit. Similar effects were seen as with the phantom and volunteer studies using external tablets; there was little effect of the Ca tablet on L1-4 BMD either with the tablet overlying bone or in the soft-tissue field. We conclude that undissolved Ca tablets introduce small artifacts on L1-L4 BMD insufficient to alter diagnostic categorization of patients. However, if only two or three vertebral levels are available, misclassification of the patients may occur due to tablet artifact. The precision of monitoring BMD over time could also be adversely affected by tablet artifact if a tablet was directly overlying bone and undetected, especially with smaller regions of interest and with lower baseline BMD.

摘要

在进行双能X线吸收法(DXA)扫描时,胃肠道中的未溶解钙片可能会影响腰椎骨矿物质密度(BMD)测定的准确性。我们对模型和志愿者进行了研究,以确定覆盖在骨骼上的钙片、软组织区域中的钙片以及同时重叠骨骼和软组织的钙片所产生的影响。对于L1 - 4,将一片500毫克的钙片粘贴在模型表面或志愿者皮肤上,导致测量的BMD平均增加≤2.2%(低于腰椎最小显著变化),对L1 - 4的T值改变很小。然而,一片覆盖在上面的钙片对单个椎体的BMD有显著影响。钙片含钙量越高、BMD越低,影响越大,在体内系列研究中单个椎体BMD平均增加高达12.6%。DXA软件的最新版本成功地将软组织区域中的钙片识别为伪影,并在分析中排除了钙片像素。然而,重叠骨骼和软组织的钙片需要操作人员进行干预,要么排除受影响的椎体,要么通过将钙片映射为伪影和/或中性区域来调整边缘。一名志愿者吞咽的石蜡包裹的钙片在胃肠道转运过程中通过DXA进行了连续追踪。观察到的效果与使用外部钙片的模型和志愿者研究相似;无论是钙片覆盖在骨骼上还是在软组织区域,钙片对L1 - 4 BMD的影响都很小。我们得出结论,未溶解的钙片会在L1 - L4 BMD上产生小的伪影,不足以改变患者的诊断分类。然而,如果只有两三个椎体水平可用,由于钙片伪影可能会导致患者分类错误。如果一片钙片直接覆盖在骨骼上且未被检测到,尤其是在感兴趣区域较小且基线BMD较低的情况下,随着时间推移监测BMD的精度也可能受到钙片伪影的不利影响。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验