[Application of a jaw motion tracking device that measures six degrees of freedom using optoelectronic].

作者信息

Ogawa Takumi, Shigeta Yuko, Ando Eriko, Hirai Shinya, Suma Mayumi, Hirabayashi Rio, Ikawa Tomoko, Hosoda Yutaka, Araki Jirou, Itoh Kousuke, Kamei Shu, Fukushima Shunji, Moriyama Takeshi, Saito Iori, Kumeda Harunobu

机构信息

The Second Department of Prosthetic Dentistry, Tsurumi University School of Dentistry.

出版信息

Nihon Hotetsu Shika Gakkai Zasshi. 2006 Apr;50(2):210-8. doi: 10.2186/jjps.50.210.

Abstract

PURPOSE

A jaw motion tracking device that measures six degrees of freedom has recently been developed. Understanding jaw motion is useful, but previous measurement methods were impractical for use in dental clinics. The overall aim of this study was to demonstrate the simple operation, low cost, and high precision of a recently developed jaw tracking device. In addition, this study explored its potential clinical applications.

METHODS

In this study, we compared two jaw motion tracking devices: a digital system type of device and an optical type of device. First we established a baseline occlusal plane from which to measure jaw motion in the same subject with both devices. The jaw motion signals were sampled at a frequency of 100 Hz. The subjects were three healthy women (mean age +/- SD = 26.3 +/- 1.2 years) who were recruited from among the crown-and-bridge faculty of Tsurumi University of Dentistry. The jaw motions measured were open-close movement, sagittal border movement, and frontal border movement. In addition, the kinematic axis point was calculated from the sagittal border movement. Data from the digital system type of device and data from the optical device were compared. The data were selected to measure rotation and translocation, i.e. jaw position about protrusion, both laterotrusion and maximal opening of the mouth.

RESULTS

The root mean square (RMS) error of position measurement was 0.163 mm with MM-JI-E and 0.178 mm with the optical type of device. The RMS error of jaw motion measurement with the optical type of device was maximum at 0.8mm and minimum at 0.1mm. This was similar to that with digital system type of jaw motion tracking device.

CONCLUSION

This study showed the possibility of developing clinical applications for this jaw motion device.