Brown David M, Beardsley Robert M, Fish Richard H, Wong Tien P, Kim Rosa Y
The Methodist Hospital, Houston, Texas 77030, USA.
Retina. 2006 Jul-Aug;26(6):645-9. doi: 10.1097/01.iae.0000236473.63819.f4.
The historical extrusion/explantation rate of silicone sponge scleral buckle exoplants is 3.5 to 24.4%. This contrasts with the published 0.6 to 1.2% explantation rate of solid silicone elements. Previously reported silicone sponge exoplants studies were noncircumferential (quadrantic or multiple). This study was undertaken to assess the long-term stability of 360 degrees circumferentially placed 3 x 5 mm silicone sponge exoplants.
Interventional case series of 840 consecutive circumferentially placed 3 x 5 mm silicone sponge exoplants. A retrospective review of operative reports and patient charts was performed.
A total of 552 patients had documented follow-up over 1 year. Median follow-up was 32.8 months. Six patients underwent removal of the scleral buckling element (two for diplopia, one acute postoperative infection, and three for extrusion). Median time to removal was 7.3 months. Extenuating circumstances contributed to two of the three extrusions. Of the entire series, fewer than 1% of patients required removal of the silicone sponge implant. Excluding patients with less than 1 year of documented follow-up, the long-term infection/extrusion rate was 0.7%.
Circumferentially placed silicone scleral buckling sponge elements are very stable and are well tolerated. The explantation/extrusion rate is comparable to published solid silicone element series.
硅胶海绵巩膜扣外植体的历史性挤出/取出率为3.5%至24.4%。这与已公布的固体硅胶元件0.6%至1.2%的取出率形成对比。先前报道的硅胶海绵外植体研究并非全周性的(象限性或多个象限)。本研究旨在评估360度全周放置的3×5毫米硅胶海绵外植体的长期稳定性。
对840例连续全周放置的3×5毫米硅胶海绵外植体进行介入性病例系列研究。对手术报告和患者病历进行回顾性分析。
共有552例患者有超过1年的记录随访。中位随访时间为32.8个月。6例患者取出了巩膜扣元件(2例因复视,1例术后急性感染,3例因挤出)。取出的中位时间为7.3个月。特殊情况导致了3例挤出中的2例。在整个系列中,不到1%的患者需要取出硅胶海绵植入物。排除记录随访时间少于1年的患者,长期感染/挤出率为0.7%。
全周放置的硅胶巩膜扣海绵元件非常稳定,耐受性良好。取出/挤出率与已公布的固体硅胶元件系列相当。