Charles Alasdair E, Ness Malcolm G
School of Chemical Engineering and Advanced Materials, University of Newcastle, Newcastle upon Tyne and Croft Veterinary Hospital, Blyth, UK.
Vet Surg. 2006 Jul;35(5):438-44. doi: 10.1111/j.1532-950X.2006.00172.x.
To describe corrosion discovered on implants recovered from dogs after tibial plateau leveling osteotomy (TPLO).
Descriptive.
Dogs (n=4) that had TPLO (n=7) approximately 3 years earlier.
Implants were recovered surgically and lightly cleaned before visual inspection, scanning electron microscopy (SEM) and energy-dispersive X-ray analysis (EDX).
Multiple corrosion sites were found on the recesses of each plate; invariably, there was a corroded area present on that part of the bone screw which had been in contact with a corrosion site on a plate. Most corroded regions featured attached, nonmetallic material believed to be a precipitate arising from interaction with the local environment during electrochemical corrosion. SEM examination of TPLO plates revealed surface porosity, particularly visible in the screw hole recesses.
Surface irregularities and porosity of TPLO plates--a consequence of the casting process--are initiation sites for corrosion which, in concert with the crevice geometry between the plate and screw, result in the observed corrosion defects.
Implants used for TPLO, a commonly performed veterinary orthopedic procedures may corrode generating a number of chemical species that have the potential to cause disease.
描述在胫骨平台水平截骨术(TPLO)后从犬只体内取出的植入物上发现的腐蚀情况。
描述性研究。
约3年前接受过TPLO手术(n = 7)的犬只(n = 4)。
通过手术取出植入物,在进行目视检查、扫描电子显微镜(SEM)和能量色散X射线分析(EDX)之前进行轻度清洁。
在每个钢板的凹槽处发现了多个腐蚀部位;在与钢板上的腐蚀部位接触的骨螺钉部分,始终存在一个腐蚀区域。大多数腐蚀区域都附着有非金属材料,据信这是电化学腐蚀过程中与局部环境相互作用产生的沉淀物。对TPLO钢板的SEM检查显示出表面孔隙,在螺钉孔凹槽中尤为明显。
TPLO钢板的表面不规则和孔隙——铸造过程的结果——是腐蚀的起始部位,与钢板和螺钉之间的缝隙几何形状共同导致了观察到的腐蚀缺陷。
用于TPLO(一种常见的兽医骨科手术)的植入物可能会腐蚀,产生多种有可能引发疾病的化学物质。