Losert Heidrun, Risdal Martin, Sterz Fritz, Nysaether Jon, Köhler Klemens, Eftestøl Trygve, Wandaller Cosima, Myklebust Helge, Uray Thomas, Sodeck Gottfried, Laggner Anton N
Department of Emergency Medicine, Medical University of Vienna, Austria.
Crit Care Med. 2006 Sep;34(9):2399-405. doi: 10.1097/01.CCM.0000235666.40378.60.
Monitoring of ventilation performance during cardiopulmonary resuscitation would be desirable to improve the quality of cardiopulmonary resuscitation. To investigate the potential for measuring ventilation rate and inspiration time, we calculated the correlation in waveform between transthoracic impedance measured via defibrillator pads and tidal volume given by a ventilator.
Clinical study.
Emergency department of a tertiary care university hospital.
A convenience sample of mechanical ventilated patients (n = 32), cardiac arrest patients (n = 20), and patients after restoration of spontaneous circulation (n = 31) older than 18 were eligible.
The Heartstart 4000SP defibrillator (Laerdal Medical Cooperation, Stavanger, Norway) with additional capabilities of recording thoracic impedance changes was used.
The relationship between impedance change and tidal volume (impedance coefficient) was calculated. The mean (sd) correlations between the impedance waveform and the tidal volume waveform in the patient groups studied were .971 (.027), .969 (.032), and .967 (.035), respectively. The mean (sd) impedance coefficient for all patients in the study was .00194 (.0078) Omega/mL, and the mean (sd) specific (weight-corrected) impedance coefficient was .152 (.048) Omega/kg/mL. The measured thorax impedance change for different tidal volumes (400-1000 mL) was approximately linear.
The impedance sensor of a defibrillator is accurate in identifying tidal volumes, when chest compressions are interrupted. This also allows quantifying ventilation rates and inspiration times. However this technology, at its present state, provides only limited practical means for exact tidal volume estimation.
在心肺复苏期间监测通气性能有助于提高心肺复苏的质量。为了研究测量通气率和吸气时间的可能性,我们计算了通过除颤器电极片测得的经胸阻抗与呼吸机给予的潮气量之间波形的相关性。
临床研究。
一所三级护理大学医院的急诊科。
选取年龄大于18岁的机械通气患者(n = 32)、心脏骤停患者(n = 20)和自主循环恢复后的患者(n = 31)作为便利样本。
使用具备记录胸阻抗变化额外功能的Heartstart 4000SP除颤器(挪威斯塔万格的Laerdal Medical Cooperation公司)。
计算阻抗变化与潮气量之间的关系(阻抗系数)。在所研究的患者组中,阻抗波形与潮气量波形之间的平均(标准差)相关性分别为0.971(0.027)、0.969(0.032)和0.967(0.035)。研究中所有患者的平均(标准差)阻抗系数为0.00194(0.0078)Ω/mL,平均(标准差)比(体重校正)阻抗系数为0.152(0.048)Ω/kg/mL。不同潮气量(400 - 1000 mL)下测得的胸部阻抗变化大致呈线性。
当胸外按压中断时,除颤器的阻抗传感器在识别潮气量方面是准确的。这也使得通气率和吸气时间的量化成为可能。然而,这项技术在当前状态下,仅为精确潮气量估计提供了有限的实用方法。