Nygaard Kirsten, Odland Odd Harald, Muren Ludvig Paul
Department of Oncology and Medical Physics, Haukeland University Hospital, Bergen, Norway.
Med Dosim. 2006 Fall;31(3):201-8. doi: 10.1016/j.meddos.2006.04.002.
The Beam Imaging System 2G (BIS-2G) from Scanditronix-Wellhöfer is a two-dimensional (2D) charge-coupled device (CCD)-camera that measures the scintillation light produced by incident radiation. We examined the performance of the BIS-2G as a tool in quality control of patient boluses. In an attempt to simplify the production of the patient boluses, bolus edges were built as staircases and the dose distributions were measured and compared to the dose profiles below corresponding sloped bolus edges. Perspex plates covering half the irradiated field were used as generalized bolus edges. All BIS-2G measurements were performed using buildup of solid water while a diode measured corresponding dose profiles in a water phantom. Below the patient boluses, regions with doses < 95% and > 107% of the prescribed dose were defined. Below the edge, the relative doses measured by the BIS-2G were generally within 3% in dose and 3 mm in position compared to the diode measurements. Close to the field edge below the bolus, the BIS-2G measurements were in some cases as much as 7% lower in dose than the diode measurements. The BIS-2G measurements revealed hotspots below the patient boluses covering 1-16% of the total irradiated area. The highest point dose measured below the patient boluses ranged from 105% to 125% of the prescribed dose. For all bolus thicknesses, each edge in the staircase bolus caused a fluctuation in dose and increased the maximum dose compared to the sloped edge. For several cases, the maximum dose increased with 13% in relative dose, e.g., from 103% to 116%. The BIS-2G was found to be a useful tool in quality control of patient boluses, revealing large hot spots in the treatment volume for several patients. Bolus edges built as staircases cause considerable dose fluctuations and increase the maximum dose, and can therefore not be recommended.
来自Scanditronix-Wellhöfer公司的束流成像系统2G(BIS-2G)是一种二维(2D)电荷耦合器件(CCD)相机,用于测量入射辐射产生的闪烁光。我们研究了BIS-2G作为患者 bolus质量控制工具的性能。为了简化患者 bolus的生产,将 bolus边缘构建为阶梯状,并测量剂量分布,并与相应倾斜 bolus边缘下方的剂量剖面进行比较。覆盖一半照射野的有机玻璃平板用作广义 bolus边缘。所有BIS-2G测量均使用固体水等效物进行,同时二极管在水体模中测量相应的剂量剖面。在患者 bolus下方,定义了剂量低于规定剂量95%和高于107%的区域。在边缘下方,与二极管测量相比,BIS-2G测量的相对剂量一般在剂量上相差3%以内,位置相差3 mm以内。在 bolus下方靠近野边缘处,在某些情况下,BIS-2G测量的剂量比二极管测量低多达7%。BIS-2G测量显示患者 bolus下方存在热点,覆盖总照射面积的1-16%。在患者 bolus下方测量的最高点剂量范围为规定剂量的105%至125%。对于所有 bolus厚度,阶梯状 bolus中的每个边缘都会导致剂量波动,并与倾斜边缘相比增加最大剂量。在几种情况下,最大剂量的相对剂量增加了13%,例如从103%增加到116%。发现BIS-2G是患者 bolus质量控制中的一种有用工具,揭示了多名患者治疗体积中的大热点。构建为阶梯状的 bolus边缘会导致相当大的剂量波动并增加最大剂量,因此不推荐使用。