Downscatter correction and choice of collimator in 123I imaging.

作者信息

Small A D, Prosser J, Motherwell D W, McCurrach G M, Fletcher A M, Martin W

机构信息

Department of Medical Cardiology, Royal Infirmary, Glasgow, UK.

出版信息

Phys Med Biol. 2006 Sep 7;51(17):N307-11. doi: 10.1088/0031-9155/51/17/N03. Epub 2006 Aug 8.

Abstract

(123)I imaging is increasingly used in nuclear medicine but downscatter from high (>300 keV) energy emissions degrades the image and introduces variation in sensitivity with depth when imaging with a low energy collimator. A dual windowing technique using a matched window immediately above the principle window to correct for the effects of downscatter is described. The technique is shown to correct for the variation in sensitivity with depth and to reduce outlying scatter. Quantitative imaging of (123)I using a low energy collimator should always employ downscatter correction.

摘要

碘-123成像在核医学中的应用日益广泛,但在使用低能准直器成像时,来自高(>300keV)能量发射的散射会降低图像质量,并引入灵敏度随深度的变化。本文描述了一种双窗技术,该技术使用紧挨着主窗上方的匹配窗来校正散射的影响。结果表明,该技术可校正灵敏度随深度的变化,并减少异常散射。使用低能准直器进行碘-123定量成像时应始终采用散射校正。