Barkhaus P E, Periquet M I, Nandedkar S D
Milwaukee Veterans Administration Medical Center & Medical College of Wisconsin, Milwaukee, Wisconsin, USA.
Electromyogr Clin Neurophysiol. 2006 Jul-Aug;46(4):235-9.
The recording characteristics of surface EMG electrodes were investigated. Compound muscle action potential (CMAP) and surface recorded motor unit action potentials were recorded from different muscles, using different surface electrode shapes and sizes. The CMAP was smaller for larger surface electrodes. This was more pronounced in smaller muscles. The CMAP was minimally affected by the geometry of the recording surface. With larger surface electrodes, shunting contributes to the reduction in MUAP amplitude. This is offset by a larger uptake area which gives a much smaller reduction in the CMAP amplitude for the larger muscles.
研究了表面肌电图电极的记录特性。使用不同形状和尺寸的表面电极,从不同肌肉记录复合肌肉动作电位(CMAP)和表面记录的运动单位动作电位。较大的表面电极记录的CMAP较小。这在较小的肌肉中更为明显。CMAP受记录表面几何形状的影响最小。对于较大的表面电极,分流会导致运动单位动作电位(MUAP)幅度降低。而更大的采集面积会抵消这一影响,使得较大肌肉的CMAP幅度降低得少得多。