Accurate measurement of physical activity patterns can be used to identify sedentary behaviour and may facilitate interventions aimed at reducing inactivity.
To evaluate the activPAL physical activity monitor as a measure of posture and motion in everyday activities using observational analysis as the criterion standard.
Wearing three activPAL monitors, 10 healthy participants performed a range of randomly assigned everyday tasks incorporating walking, standing and sitting. Each trial was captured on a digital camera and the recordings were synchronised with the activPAL. The time spent in different postures was visually classified and this was compared with the activPAL output.
Intraclass correlation coefficients (ICC 2,1) for interdevice reliability ranged from 0.79 to 0.99. Using the Bland and Altman method, the mean percentage difference between the activPAL monitor and observation for total time spent sitting was 0.19% (limits of agreement -0.68% to 1.06%) and for total time spent upright was -0.27% (limits of agreement -1.38% to 0.84%). The mean difference for total time spent standing was 1.4% (limits of agreement -6.2% to 9.1%) and for total time spent walking was -2.0% (limits of agreement -16.1% to 12.1%). A second-by-second analysis between observer and monitor found an overall agreement of 95.9%.
The activPAL activity monitor is a valid and reliable measure of posture and motion during everyday physical activities.
准确测量身体活动模式可用于识别久坐行为,并可能有助于开展旨在减少不活动状态的干预措施。
以观察性分析作为标准参照,评估activPAL身体活动监测仪作为日常活动中姿势和运动测量工具的效果。
10名健康参与者佩戴3个activPAL监测仪,执行一系列随机分配的日常任务,包括行走、站立和坐下。每次试验均用数码相机记录,并将录像与activPAL监测仪同步。通过视觉分类确定在不同姿势下所花费的时间,并将其与activPAL监测仪的输出结果进行比较。
设备间可靠性的组内相关系数(ICC 2,1)范围为0.79至0.99。采用Bland和Altman方法,activPAL监测仪与观察结果之间,就坐总时间的平均百分比差异为0.19%(一致性界限为-0.68%至1.06%),直立总时间的平均百分比差异为-0.27%(一致性界限为-1.38%至0.84%)。站立总时间的平均差异为1.4%(一致性界限为-6.2%至9.1%),行走总时间的平均差异为-2.0%(一致性界限为-16.1%至12.1%)。观察者与监测仪之间的逐秒分析显示总体一致性为95.9%。
activPAL活动监测仪是日常身体活动中姿势和运动的有效且可靠的测量工具。