To investigate the causal relationship between acute postoperative endophthalmitis (POE) after cataract surgery and the biomaterial properties of the intraocular lens (IOLs) implanted.
This retrospective cohort study included all patients who had undergone cataract surgery with IOL implantation at the Lyon Croix-Rousse University Hospital between 1st January 1994 and 31st December 2004. Details respecting the type of IOL implanted (material and manufacturer) were meticulously recorded. The number of patients presenting with POE within 6 weeks of cataract surgery was documented together with their medical characteristics. These data were then compared, and Fisher's exact test was used to establish the significance of any apparent associations.
Eight of the 5837 eyes manifested acute POE (0.14%). Seven of these were composed of polymethylmethacrylate (PMMA) and one of heparinized PMMA. Patients with PMMA IOLs carried a higher risk of developing POE than did those implanted with either heparinized PMMA (P=0.001), hydrophilic acrylic, or hydrophobic acrylic IOLs (P=0.002).
The incidence of acute POE after cataract surgery in our hospital is similar to that currently reported for other institutions in developed countries. Our results add further evidence that IOL material and type are factors contributing to the risk to develop an acute POE after cataract surgery, and that PMMA IOLs may be associated with an increased risk of POE.
研究白内障手术后急性眼内炎(POE)与所植入人工晶状体(IOL)生物材料特性之间的因果关系。
这项回顾性队列研究纳入了1994年1月1日至2004年12月31日期间在里昂十字-鲁西大学医院接受白内障手术并植入IOL的所有患者。详细记录所植入IOL的类型(材料和制造商)。记录白内障手术后6周内出现POE的患者数量及其医学特征。然后对这些数据进行比较,并使用Fisher精确检验确定任何明显关联的显著性。
5837只眼中有8只出现急性POE(0.14%)。其中7只为聚甲基丙烯酸甲酯(PMMA)材质,1只为肝素化PMMA材质。植入PMMA IOL的患者发生POE的风险高于植入肝素化PMMA(P=0.001)、亲水性丙烯酸或疏水性丙烯酸IOL的患者(P=0.002)。
我院白内障手术后急性POE的发生率与目前发达国家其他机构报告的发生率相似。我们的结果进一步证明,IOL材料和类型是白内障手术后发生急性POE风险的影响因素,且PMMA IOL可能与POE风险增加有关。