Steen Jensen J, Nilsen R, Zeffer J, Fisk J, Hartz C
ISPO, Copenhagen, Denmark.
Prosthet Orthot Int. 2006 Aug;30(2):195-212. doi: 10.1080/03093640600794692.
In a prospective, controlled study, 186 prosthetic rubber feet of different designs were fitted to amputees with trans-tibial prostheses. There were 158 amputees available for follow-up. The amputee compliance was good and 89% were satisfied with their device. After 18 months of use one VI-Solid rubber foot from VVAF in Cambodia had failed, as compared to 11% failures with the same foot with a heel cavity, 3% with the EB-1 sandwich construction from POF in Vietnam, all performing significantly better than the 62% encountered with the vulcanized rubber foot from ICRC in Cambodia; the latter representing half of the feet failing in amputees walking bare-footed. Nearly all failures were located at the foot-sole or the keel. The failure rate was 20% for the HI-Cambodia foot after 12 months. After 24 months of use, still only one VI-Solid rubber foot had failed and this foot design was found superior to the others in regard of durability.
在一项前瞻性对照研究中,为186名佩戴经胫假肢的截肢者安装了不同设计的假肢橡胶脚。有158名截肢者可供随访。截肢者的依从性良好,89%的人对他们的假肢满意。使用18个月后,柬埔寨VVAF公司的一款VI-Solid橡胶脚出现了故障,相比之下,带有后跟腔的同款脚的故障率为11%,越南POF公司的EB-1夹层结构脚的故障率为3%,所有这些都明显优于柬埔寨红十字国际委员会的硫化橡胶脚62%的故障率;后者占赤足行走截肢者中出现故障的脚的一半。几乎所有故障都发生在鞋底或龙骨处。HI-柬埔寨脚在使用12个月后的故障率为20%。使用24个月后,仍然只有一款VI-Solid橡胶脚出现故障,并且发现这种脚的设计在耐用性方面优于其他脚。