一种用于检测人血浆中17α-乙炔雌二醇的竞争性蛋白结合放射分析法。

A competitive protein binding radioassay for 17alpha-ethynylestradiol in human plasma.

作者信息

Verma P, Curry C, Crocker C, Titus-Dillon P, Ahluwalia B

出版信息

Clin Chim Acta. 1975 Sep 16;63(3):363-8. doi: 10.1016/0009-8981(75)90058-3.

Abstract

A competitive protein binding radioassay system using a rabbit uterine cytosol preparation for the measurement of ethynylestradiol [EE2e] in human plasma has been developed. Sephadex LH-20 column chromatography is used for the separation of the estrogens prior to assay. EE2 levels were measured in 23 normotensive subjects on contraceptive pills containing either 50 mug of EE2 or mestranol. The EE2 values ranged from 144 to 248 pg with a mean of 211 pg (S.D. +/- 37.94). The method is sensitive enough to detect 25 pg of EE2 in plasma. The intra- and inter-assay coefficients of variation were 7.54% and 10.75%, respectively. Water blanks and plasma blanks from subjects not on pills gave negligible readings for EE2.

摘要

已开发出一种竞争性蛋白质结合放射分析系统,该系统使用兔子宫胞质溶胶制剂来测量人血浆中的乙炔雌二醇[EE2e]。在测定之前,使用葡聚糖凝胶LH - 20柱色谱法分离雌激素。对23名服用含50μg EE2或炔雌醇避孕药的血压正常受试者的EE2水平进行了测量。EE2值范围为144至248 pg,平均值为211 pg(标准差±37.94)。该方法灵敏度足以检测出血浆中25 pg的EE2。测定内和测定间变异系数分别为7.54%和10.75%。未服用避孕药的受试者的水空白和血浆空白对EE2的读数可忽略不计。