Laymon Charles M, Turkington Timothy G
Department of Radiology, University of Pittsburgh, UPMC, Pittsburgh, PA 15213, USA.
Nucl Med Commun. 2006 Nov;27(11):901-9. doi: 10.1097/01.mnm.0000239478.46157.64.
Single photon emission computed tomography (SPECT) with 511 keV photons is a challenging modality and collimators for this purpose require trade-offs among resolution, sensitivity and septal penetration. While PET is the modality of choice for imaging at 511 keV, there are some procedures, e.g., dual-isotope imaging, in which 511 keV SPECT has a role.
To measure the imaging properties of a VPC-93 SPECT collimator designed for imaging at 511 keV and to isolate the effects of septal penetration.
NaI gamma camera projection images of (18)F (511 keV) and (99m)Tc (140 keV) point sources were measured and the corresponding modulation transfer functions calculated. The projection images were reconstructed via filtered back-projection to obtain the tomographic three-dimensional (3-D) point spread function. Differences between the 511 and 140 keV results were attributed mainly to septal penetration. Contrast measurements were made separately using (18)F and (99m)Tc of a 20 cm phantom containing hot spheres and a warm background. Both isotopes were also used in imaging studies of a 3-D Hoffman brain phantom.
Reconstructed 511 keV point source images were spatially extended with more than half of the total reconstructed counts appearing away from the point source region. The number of false counts contained in the image as a function of distance from the true source location remains approximately constant for large distances out to at least 14 cm. Septal penetration results in a rapid roll-off with spatial frequency of collimator response. The response of the collimator to 511 keV photons falls to half of its 0-frequency response at 0.03 cm(-1). For 140 keV photons this value is 0.20 cm(-1). A result is reduced image contrast as measured in the phantom sphere studies. Septal penetration causes image degradation through large-scale blurring. Image noise characteristics are modified and correlations are extended into many transaxial planes.
Both 2-D and 3-D point spread functions for 511 and 140 keV photons using the VPC-93 collimator have been measured. Septal penetration unfavourably affects image resolution and changes image noise characteristics. Without compensation, the effects of septal penetration are readily apparent in images of real objects.
利用511千电子伏特光子的单光子发射计算机断层扫描(SPECT)是一种具有挑战性的成像方式,用于此目的的准直器需要在分辨率、灵敏度和隔板穿透之间进行权衡。虽然正电子发射断层扫描(PET)是511千电子伏特成像的首选方式,但在一些程序中,例如双同位素成像,511千电子伏特SPECT也能发挥作用。
测量一款为511千电子伏特成像设计的VPC - 93 SPECT准直器的成像特性,并分离出隔板穿透的影响。
测量了(18)F(511千电子伏特)和(99m)Tc(140千电子伏特)点源的碘化钠γ相机投影图像,并计算了相应的调制传递函数。通过滤波反投影重建投影图像以获得断层三维(3 - D)点扩散函数。511千电子伏特和140千电子伏特结果之间的差异主要归因于隔板穿透。使用含有热球和温暖背景的20厘米体模,分别用(18)F和(99m)Tc进行对比度测量。这两种同位素也用于三维霍夫曼脑体模的成像研究。
重建的511千电子伏特点源图像在空间上扩展,超过一半的总重建计数出现在远离点源区域的地方。图像中包含的虚假计数数量作为与真实源位置距离的函数,在至少14厘米的大距离范围内大致保持恒定。隔板穿透导致准直器响应的空间频率快速下降。准直器对511千电子伏特光子的响应在0.03厘米^(-1)时降至其零频率响应的一半。对于140千电子伏特光子,该值为0.20厘米^(-1)。结果是在体模球体研究中测量的图像对比度降低。隔板穿透通过大规模模糊导致图像退化。图像噪声特性被改变,相关性扩展到许多横断面。
使用VPC - 93准直器测量了511千电子伏特和140千电子伏特光子的二维和三维点扩散函数。隔板穿透对图像分辨率有不利影响,并改变图像噪声特性。如果不进行补偿,隔板穿透的影响在真实物体的图像中很容易显现出来。