Vogel Guy, Chechick Aharon, Pritch Perry Moshe, Brosh Tamar
The Chaim Sheba Medical Center, Ramat Gan, Israel.
Int Orthop. 2007 Dec;31(6):811-5. doi: 10.1007/s00264-006-0266-9. Epub 2006 Oct 17.
The conventional procedure for displaced fracture of the surgical neck of the humerus uses straight pins inserted from the lateral aspect towards the head of the humerus. The objective of this study was to compare the mechanical properties of fixation by contoured (curved) pins to those fixed with regular straight pins. A transverse osteotomy was made in 30 fresh-frozen sheep humeri in the proximal metaphyseal bone region, and pins were inserted using either three parallel straight pins or three contoured pins in different planes. The assemblies were subjected to bending or rotational stresses at the fracture site. Loads versus deformations were acquired during loading and rigidity was calculated. Results showed that in bending, fixation with straight pins was 31% more rigid compared to contoured pins (p < 0.001), and in torsion, fixation using contoured pins was 21% more rigid compared to straight pins (p < 0.001). A combination of the two fixation types should be considered.
治疗肱骨外科颈移位骨折的传统方法是从外侧向肱骨头插入直针。本研究的目的是比较异形(弯曲)针固定与常规直针固定的力学性能。对30根新鲜冷冻的绵羊肱骨近端干骺端区域进行横向截骨,分别在不同平面使用三根平行直针或三根异形针进行插入。对骨折部位的组件施加弯曲或旋转应力。加载过程中记录载荷与变形情况,并计算刚度。结果显示,在弯曲试验中,直针固定的刚度比异形针高31%(p<0.001);在扭转试验中,异形针固定的刚度比直针高21%(p<0.001)。应考虑两种固定方式的联合应用。