Abstract

The stiffness of a bipolar pacing lead, particularly between anode and cathode, may be responsible for myocardial penetration and perforation. Following an unprecedented 7% incidence of high threshold exit block with a single model bipolar ventricular endocardial lead, a study was undertaken to compare pacing lead stiffness between anode and cathode of six models of bipolar leads from two manufacturers; Telectronics (T) and Medtronics (M). Four leads had polyurethane insulation; T 030-284 (Laser Dish), T 329-259 (Cordis, Encor), M 4012 (Target Tip), and M 4004 (Capsure). Two leads had silicone rubber insulation; M 5026 (Capsure) and M 5024 (Capsure SP). All leads were subjected to two stiffness tests. The Tip Deflection Test involved securing the lead at 45 degrees at the indifferent electrode and applying a force to deflect the tip 5 mm. The three point bending test involved placing the lead over two fixed bars in contact with the anode and cathode. Midway a third bar was pushed onto the lead and the force to deflect the lead 2 mm was recorded. The results showed that pacing leads with polyurethane insulation were much stiffer than those with silicone rubber insulation. The T 030-284 because of its construction was found to be the stiffnest. The next stiffnest was the M 4012. Both these leads had an unacceptable incidence of high threshold exit block; 7% with the T 030-284 (89 implants) and 3% with the M 4012 (102 implants). No cases of high threshold exit block were documented with the other four pacing leads and in particular the silicone rubber M 5026 (344 implants).(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

双极起搏导线的硬度,尤其是阳极和阴极之间的硬度,可能是导致心肌穿透和穿孔的原因。在一款单型号双极心室心内膜导线出现前所未有的7%高阈值出口阻滞发生率后,开展了一项研究,比较来自两家制造商(Telectronics(T)和Medtronic(M))的六种型号双极导线阳极和阴极之间的起搏导线硬度。四种导线采用聚氨酯绝缘;T 030 - 284(激光盘)、T 329 - 259(Cordis,Encor)、M 4012(目标尖端)和M 4004(捕获)。两种导线采用硅橡胶绝缘;M 5026(捕获)和M 5024(捕获SP)。所有导线都进行了两项硬度测试。尖端偏转测试包括将导线在无关电极处固定成45度,并施加一个力使尖端偏转5毫米。三点弯曲测试包括将导线放置在与阳极和阴极接触的两根固定杆上。在中间位置将第三根杆推到导线上,并记录使导线偏转2毫米所需的力。结果表明,采用聚氨酯绝缘的起搏导线比采用硅橡胶绝缘的导线硬得多。由于其结构,T 030 - 284被发现是最硬的。其次是M 4012。这两种导线都有不可接受的高阈值出口阻滞发生率;T 030 - 284为7%(89次植入),M 4012为3%(102次植入)。其他四种起搏导线,特别是硅橡胶材质的M 5026(344次植入),未记录到高阈值出口阻滞病例。(摘要截取自250字)

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验