通用电气医疗集团eXplore VISTA双环小动物正电子发射断层扫描仪的性能评估

Performance evaluation of the GE healthcare eXplore VISTA dual-ring small-animal PET scanner.

作者信息

Wang Yuchuan, Seidel Jurgen, Tsui Benjamin M W, Vaquero Juan J, Pomper Martin G

机构信息

Department of Radiology, Johns Hopkins Medical Institutions, Baltimore, Maryland, USA.

出版信息

J Nucl Med. 2006 Nov;47(11):1891-900.

Abstract

UNLABELLED

We evaluated the performance characteristics of the eXplore VISTA dual-ring small-animal PET scanner, a stationary, ring-type, depth-of-interaction (DOI) correcting system designed to simultaneously maximize sensitivity, resolution, and resolution uniformity over a field of view sufficient to image rodent-sized animals.

METHODS

We measured the intrinsic spatial resolution response of the VISTA detector modules, spatial and volume resolution throughout a representative portion of the field of view, and imaged several common resolution phantoms to provide a qualitative picture of resolution performance. We obtained an axial sensitivity profile and measured central point source sensitivity, scatter fractions and noise equivalent count (NEC) rates for rat- and mouse-sized objects using different energy windows, and count rate linearity. In addition, we measured the energy and timing resolution of both of the crystal layers (cerium-doped gadolinium orthosilicate and cerium-doped lutetium-yttrium orthosilicate) that give VISTA machines a DOI compensation capability. We examined the effectiveness of this DOI compensation by comparing spatial resolution measurements with and without the DOI correction enabled. Finally, several animal studies were included to illustrate system performance in the field.

RESULTS

Spatial and volume resolutions averaged approximately 1.4 mm and 2.9 mm(3), respectively (with 3-dimensional Fourier rebinning and 2-dimensional filtered backprojection image reconstructions and an energy window of 250-700 keV), along the central axis of the scanner, and the spatial resolution was better than 1.7 mm and 2.1 mm at 1 and 2 cm off the central axis, respectively. Central point source sensitivity measured approximately 4% with peak NEC rates of 126.8 kcps at 455 kBq/mL and 77.1 kcps at 141 kBq/mL for mouse- and rat-sized uniform, cylindric phantoms, respectively. The radial spatial resolution at 2.8 cm off axis with DOI compensation was 2.5 mm but degraded (by 56%) to 3.9 mm without DOI compensation (as would be the case with a geometrically identical scanner without DOI correction capability).

CONCLUSION

These results indicate that the VISTA small-animal PET scanner is well suited to imaging rodent-sized animals. The combination of high spatial resolution, resolution uniformity, sensitivity, and count rate performance, made possible in part by the novel use of phoswich detector modules, confers significant technical advantages over machines with similar geometry but without DOI correction capability.

摘要

未标注

我们评估了eXplore VISTA双环小动物正电子发射断层扫描仪的性能特性,这是一种固定式环型深度交互(DOI)校正系统,旨在在足以对啮齿类动物大小的动物进行成像的视野范围内,同时最大化灵敏度、分辨率和分辨率均匀性。

方法

我们测量了VISTA探测器模块的固有空间分辨率响应、整个视野代表性部分的空间和体积分辨率,并对几个常见分辨率体模进行成像,以提供分辨率性能的定性图像。我们获得了轴向灵敏度曲线,并使用不同的能量窗测量了大鼠和小鼠大小物体的中心点源灵敏度、散射分数和噪声等效计数(NEC)率,以及计数率线性。此外,我们测量了赋予VISTA机器DOI补偿能力的两个晶体层(掺铈的正硅酸钆和掺铈的镥钇正硅酸盐)的能量和时间分辨率。我们通过比较启用和未启用DOI校正时的空间分辨率测量结果,研究了这种DOI补偿的有效性。最后,纳入了几项动物研究以说明该系统在实际应用中的性能。

结果

沿扫描仪中心轴,空间分辨率和体积分辨率平均分别约为1.4毫米和2.9立方毫米(采用三维傅里叶重新组合和二维滤波反投影图像重建,能量窗为250 - 700 keV),在离中心轴1厘米和2厘米处,空间分辨率分别优于1.7毫米和2.1毫米。对于小鼠和大鼠大小的均匀圆柱形体模,中心点源灵敏度约为4%,在455 kBq/mL时峰值NEC率为126.8 kcps,在141 kBq/mL时为77.1 kcps。在离轴2.8厘米处,有DOI补偿时的径向空间分辨率为2.5毫米,但无DOI补偿时降至3.9毫米(下降56%)(与几何结构相同但无DOI校正能力的扫描仪情况相同)。

结论

这些结果表明,VISTA小动物正电子发射断层扫描仪非常适合对啮齿类动物大小的动物进行成像。高空间分辨率、分辨率均匀性、灵敏度和计数率性能的结合,部分得益于磷光体探测器模块的新颖应用,与具有类似几何结构但无DOI校正能力的机器相比,具有显著的技术优势。