高分辨率且经济高效的台式临床前PET扫描仪easyPET.3D的首次性能评估。

First performance evaluation of easyPET.3D, a high-resolution and cost-effective benchtop preclinical PET scanner.

作者信息

Ribeiro Fabiana M, Encarnação Pedro M C C, Silva Ana L M, Correia Pedro M M, Pinto Afonso X, Oliveira Regina G, Sereno José, Pais Mariana Lapo, Abrunhosa Antero, Castro Ismael F, Santos Ana C, Veloso João F C A

机构信息

Institute for Nanostructures, Nanomodelling and Nanofabrication (i3N), Physics Department (DFis), University of Aveiro (UA), Aveiro, 3810-193, Portugal.

Institute for Nuclear Sciences Applied to Health (ICNAS), University of Coimbra, Coimbra, 3000-548, Portugal.

出版信息

EJNMMI Phys. 2025 May 26;12(1):48. doi: 10.1186/s40658-025-00757-0.

Abstract

PURPOSE

EasyPET.3D is a preclinical positron emission tomography (PET) scanner with a unique scanning method based on two face-to-face detector modules with two axes of motion. Its performance evaluation is presented using the NEMA NU-4 standards and an animal model.

METHODS

Each detector module consists of a 32 × 2 pixelated cerium-doped lutetium-yttrium oxyorthosilicate (LYSO:Ce) scintillator array with individual crystals of 2.0 × 2.0 × 30 mm. The crystal arrays are coupled to 1.3 × 1.3 mm silicon photomultipliers (SiPMs). The transaxial field-of-view (FoV) is adjustable up to 48 mm in diameter, and the axial length is 73 mm. The performance characterization includes spatial resolution, sensitivity, count rate, scatter fraction (SF), and image quality (IQ) measurements. Furthermore, mice experiments were conducted to evaluate the in vivo imaging capability.

RESULTS

Spatial resolution at the FoV centre (CFoV) in radial, tangential, and axial directions was 0.98±0.08, 0.97±0.06 and 0.94±0.08 mm, respectively. An absolute sensitivity of 0.23% was measured at CFoV. The mouse-like phantom SF was 16% (913 cps at 18 MBq). Recovery coefficients in the IQ phantom varied from 21±34% to 85±50% (1 to 5 mm diameter rods, accordingly). The uniformity was 17.6%, and spill-over ratios for water-filled and air-filled chambers were 0.49 and 0.40, respectively.

CONCLUSION

EasyPET.3D geometry favours the reduction of parallax error, despite its low sensitivity. The system linearity is suitable for the low range of activities (7-8 MBq) used for mice imaging. The overall performance showed that easyPET.3D has potential for entry-level preclinical applications.

摘要

目的

EasyPET.3D是一款临床前正电子发射断层扫描(PET)扫描仪,采用独特的扫描方法,基于两个面对面的探测器模块和两个运动轴。使用NEMA NU-4标准和动物模型对其性能进行评估。

方法

每个探测器模块由一个32×2像素化的掺铈硅酸镥钇(LYSO:Ce)闪烁体阵列组成,单个晶体尺寸为2.0×2.0×30毫米。晶体阵列与1.3×1.3毫米的硅光电倍增管(SiPM)耦合。横向视野(FoV)直径最大可调节至48毫米,轴向长度为73毫米。性能表征包括空间分辨率、灵敏度、计数率、散射分数(SF)和图像质量(IQ)测量。此外,还进行了小鼠实验以评估其体内成像能力。

结果

在视野中心(CFoV)处,径向、切向和轴向的空间分辨率分别为0.98±0.08、0.97±0.06和0.94±0.08毫米。在CFoV处测得的绝对灵敏度为0.23%。类小鼠体模的SF为16%(18 MBq时为913 cps)。IQ体模中的恢复系数在21±34%至85±50%之间变化(对应直径为1至5毫米的棒)。均匀性为17.6%,充水腔和充气腔的溢出率分别为0.49和0.40。

结论

尽管EasyPET.3D灵敏度较低,但其几何结构有利于减少视差误差。该系统线性度适用于小鼠成像所用的低活度范围(7 - 8 MBq)。整体性能表明,EasyPET.3D在入门级临床前应用方面具有潜力。