Meyhoefer Juergen, Lehmann Heiko, Minden Hans-Heinrich, Butter Christian
Department of Cardiology, Heart Center Brandenburg in Bernau, Ladeburger Strasse 17, 16321 Bernau, Germany.
Europace. 2006 Dec;8(12):1070-2. doi: 10.1093/europace/eul129. Epub 2006 Nov 10.
We describe the closure of a subclavian artery puncture site with a percutaneous suture device after removal of a pacemaker lead 1 week after its inadvertent positioning in the left ventricle via the subclavian artery. The lead was retracted from the left ventricle into the aorta and linked to a guiding catheter introduced via femoral artery access. The lead and the guiding catheter were removed from the artery to the subclavian area. This manoeuvre allowed the placement of a percutaneous arterial suture device (Perclose) to close the puncture site.
我们描述了在经锁骨下动脉意外将起搏器导线置于左心室1周后,移除起搏器导线后使用经皮缝合装置封闭锁骨下动脉穿刺部位的情况。导线从左心室回缩至主动脉,并与经股动脉途径引入的引导导管相连。导线和引导导管从动脉移除至锁骨下区域。这一操作使得能够放置经皮动脉缝合装置(Perclose)来封闭穿刺部位。