Optom Vis Sci. 2006 Nov;83(11):836-42. doi: 10.1097/01.opx.0000238722.34167.cc.

Abstract

PURPOSE

The purpose of this study was to compare the new, computerized Aniseikonia Inspector with the Space Eikonometer for the measurement of aniseikonia.

METHODS

Eighteen subjects, ages 21 to 61 years, with normal binocular vision and normal visual acuity had aniseikonia measured with both the Aniseikonia Inspector Version I and the Space Eikonometer. Aniseikonia was measured first with the subjects' habitual refractive correction and then with afocal size lenses of 1%, 2%, and 3.5% added in random order before the right and left eyes. Measurements were taken initially with the Aniseikonia Inspector and on a subsequent day with the Space Eikonometer.

RESULTS

For the Space Eikonometer, the slopes of the lines for the relationship between the measured aniseikonia and the induced magnification differences in the vertical and horizontal meridians are not significantly different from 1.0. For the Aniseikonia Inspector, the slopes of the lines in the vertical and the horizontal meridians are less than and significantly different from 1.0. On average, the Aniseikonia Inspector underestimates the magnitude of induced aniseikonia, predicting only 68% and 61% of the overall size lens magnification in the vertical and horizontal meridians, respectively. The corresponding values of the Space Eikonometer in the vertical and horizontal meridians are 99% and 93%. Variability is greater with the Space Eikonometer than the Aniseikonia Inspector.

CONCLUSION

The Space Eikonometer appears to be measuring the induced aniseikonia appropriately, whereas the Aniseikonia Inspector underestimates the amount of aniseikonia. However, the Space Eikonometer shows greater measurement variability. Modification of the Aniseikonia Inspector or the testing conditions should be pursued in future studies.

摘要

目的

本研究旨在比较新型计算机化的不等像检查仪与空间像仪在测量不等像方面的性能。

方法

选取18名年龄在21至61岁之间、双眼视力正常且视敏度正常的受试者,使用不等像检查仪I型和空间像仪测量其不等像。首先在受试者习惯的屈光矫正状态下测量不等像,然后在左右眼前随机依次添加1%、2%和3.5%的无焦镜片后再次测量。最初使用不等像检查仪进行测量,次日使用空间像仪进行测量。

结果

对于空间像仪,测量的不等像与垂直和水平子午线诱导放大率差异之间关系的直线斜率与1.0无显著差异。对于不等像检查仪,垂直和水平子午线直线的斜率均小于1.0且与1.0有显著差异。平均而言,不等像检查仪低估了诱导不等像的大小,在垂直和水平子午线中分别仅预测了整体镜片放大率的68%和61%。空间像仪在垂直和水平子午线中的相应值分别为99%和93%。空间像仪的变异性大于不等像检查仪。

结论

空间像仪似乎能适当地测量诱导不等像,而不等像检查仪低估了不等像的量。然而,空间像仪显示出更大的测量变异性。未来的研究应致力于对不等像检查仪或测试条件进行改进。