两种基于计算机的用于测量影像不等的新测试方法的评估。

The evaluation of two new computer-based tests for measurement of Aniseikonia.

作者信息

Fullard Roderick J, Rutstein Robert P, Corliss David A

机构信息

School of Optometry, University of Alabama at Birmingham, Birmingham, AL 35294-0010, USA.

出版信息

Optom Vis Sci. 2007 Dec;84(12):1093-100. doi: 10.1097/OPX.0b013e31815b9e4c.

Abstract

PURPOSE

To evaluate the accuracy and repeatability of size lens induced aniseikonia measurement with the Aniseikonia Inspector Version 1 and a newer customized version of the Aniseikonia Inspector, Version 2.

METHODS

Aniseikonia was measured on 27 subjects with both versions of the Aniseikonia Inspector in normal room illumination. Measurements of induced aniseikonia were made using size lenses in a randomized order. Twenty-five subjects were further tested in the dark using target sizes of equal visual angle for both tests. Repeatability of the intrinsic aniseikonia measurement was assessed on five subjects using randomized testing order for instrument and light and dark measurements.

RESULTS

In normal illumination, the mean slopes for plots of induced aniseikonia vs. size lens magnification for Version 1 were 0.883 and 0.838 for the vertical and horizontal meridians, respectively. For Version 2, the corresponding slopes were 1.162 and 1.043. In the dark and using targets of the same size for both tests, the slopes for Version 1 were 1.038 in the vertical meridian and 0.866 in the horizontal meridian whereas for Version 2, the slopes were 1.195 in the vertical meridian and 1.127 in the horizontal meridian. The amount of underestimation or overestimation within any given testing condition showed considerable intersubject variation. Version 1 was more repeatable than 2, particularly in the vertical meridian.

CONCLUSIONS

On average, the most accurate and repeatable measurement of aniseikonia was found with Version 1 in the vertical meridian in the dark. Measurement of aniseikonia in the horizontal meridian appears to be less reliable. Version 2 overestimates size lens-induced aniseikonia under all testing conditions. Intersubject variation in slopes of induced aniseikonia vs. size lens magnification should be further addressed.

摘要

目的

评估使用1版和更新的定制版2版的不等像视检查仪测量大小透镜诱导性不等像的准确性和可重复性。

方法

在正常室内照明条件下,使用这两个版本的不等像视检查仪对27名受试者进行不等像视测量。使用大小透镜以随机顺序进行诱导性不等像的测量。对25名受试者在黑暗环境中进一步进行测试,两次测试均使用等视角的目标大小。使用仪器、光照和黑暗测量的随机测试顺序,对5名受试者评估固有不等像测量的可重复性。

结果

在正常照明下,1版诱导性不等像与大小透镜放大倍数关系图的垂直和水平子午线的平均斜率分别为0.883和0.838。对于2版,相应的斜率分别为1.162和1.043。在黑暗环境中且两次测试使用相同大小的目标时,1版垂直子午线的斜率为1.038,水平子午线的斜率为0.866;而对于2版,垂直子午线的斜率为1.195,水平子午线的斜率为1.127。在任何给定测试条件下,低估或高估的程度在受试者之间存在相当大的差异。1版比2版更具可重复性,尤其是在垂直子午线方面。

结论

平均而言,在黑暗环境中垂直子午线方向,使用1版测量不等像最为准确和可重复。水平子午线方向的不等像测量似乎不太可靠。在所有测试条件下,2版均高估了大小透镜诱导性不等像。诱导性不等像与大小透镜放大倍数关系图斜率的受试者间差异应进一步研究。