Díaz Marlén Pérez, Rizo Oscar Díaz, Díaz Adlin López, Aparicio Eric Estévez, Díaz Reinaldo Roque
Center for Studies on Electronics & Info Tech. Images Group, Central University of Las Villas, Santa Clara 54830, Cuba.
J Zhejiang Univ Sci B. 2006 Dec;7(12):947-56. doi: 10.1631/jzus.2006.B0947.
A discriminant method for optimizing activity in nuclear medicine studies is validated by comparison with ROC (received operating characteristic)-curves. The method is tested in 21 single photon emission computerized tomography (SPECT), performed with a cardiac phantom. Three different lesions (L(1), L(2) and L(3)) were placed in the myocardium-wall by pairs for each SPECT. Three activities (84, 37 or 18.5 MBq) of 99mTc were used as background. Linear discriminant analysis was used to select the parameters that characterize image quality among the measured variables in the images [(Background-to-Lesion (B/L(i)) and Signal-to-Noise (S(i)/N) ratios)]. Two clusters with different image quality (P=0.021) were obtained. The ratios B/L(1), B/L(2) and B/L(3) are the parameters used to construct the function with 100% of cases correctly classified into the clusters. The value of 37 MBq was the lowest tested activity for which good results for the B/L(i) ratios were obtained. The result coincides with the applied ROC-analysis (r=0.89).
一种用于优化核医学研究中活性的判别方法通过与ROC(接收操作特征)曲线比较得到验证。该方法在使用心脏模型进行的21次单光子发射计算机断层扫描(SPECT)中进行了测试。每次SPECT在心肌壁上成对放置三种不同的病变(L(1)、L(2)和L(3))。使用三种活度(84、37或18.5 MBq)的99mTc作为本底。线性判别分析用于从图像中的测量变量[(本底与病变(B/L(i))和信噪比(S(i)/N))]中选择表征图像质量的参数。获得了具有不同图像质量的两个聚类(P = 0.021)。B/L(1)、B/L(2)和B/L(3)比值是用于构建函数的参数,100%的病例被正确分类到聚类中。37 MBq是获得良好B/L(i)比值测试结果的最低测试活度。该结果与应用的ROC分析结果一致(r = 0.89)。