细胞仪灵敏度的提高对检测低荧光CD5阳性B细胞的影响。

The effect of improvements in cytometer sensitivity on the detection of CD5-positive B cells with dim fluorescence.

作者信息

Givan A L, Calvert J E, Shenton B K

机构信息

Department of Surgery, Newcastle University Medical School, England.

出版信息

Cytometry. 1991;12(4):360-5. doi: 10.1002/cyto.990120411.

Abstract

When antigen density on the surface of a cell population is low and variable, the percentage of that population determined to express the antigen (i.e., to be positively stained) depends directly on the sensitivity of the flow cytometer for resolving particles which are dimly fluorescent from those which are unstained. In this study, the sensitivity of a commercial flow cytometer has been improved by changes in the photomultiplier tube, the fluorescence filter, and the amount of stray light entering the fluorescence channel. In a model system with human lymphocytes, modifications to these factors increased the percent of the B-lymphocyte population found to express the CD5 antigen.

摘要

当细胞群体表面的抗原密度较低且存在差异时,该群体中被确定表达抗原(即呈阳性染色)的百分比直接取决于流式细胞仪区分低荧光颗粒与未染色颗粒的灵敏度。在本研究中,通过改变光电倍增管、荧光滤光片以及进入荧光通道的杂散光量,提高了商用流式细胞仪的灵敏度。在一个人类淋巴细胞模型系统中,对这些因素的调整增加了发现表达CD5抗原的B淋巴细胞群体的百分比。