EPR dosimetry of radiotherapy photon beams in inhomogeneous media using alanine films.

作者信息

Osterås Bjørn Helge, Hole Eli Olaug, Olsen Dag Rune, Malinen Eirik

机构信息

Department of Radiation Biology, Institute for Cancer Research, The Norwegian Radium Hospital, Montebello, N-0310 Oslo, Norway.

出版信息

Phys Med Biol. 2006 Dec 21;51(24):6315-28. doi: 10.1088/0031-9155/51/24/003. Epub 2006 Nov 21.

Abstract

In the current work, EPR (electron paramagnetic resonance) dosimetry using alanine films (134 microm thick) was utilized for dose measurements in inhomogeneous phantoms irradiated with radiotherapy photon beams. The main phantom material was PMMA, while either Styrofoam or aluminium was introduced as an inhomogeneity. The phantoms were irradiated to a maximum dose of about 30 Gy with 6 or 15 MV photons. The performance of the alanine film dosimeters was investigated and compared to results from ion chamber dosimetry, Monte Carlo simulations and radiotherapy treatment planning calculations. It was found that the alanine film dosimeters had a linear dose response above approximately 5 Gy, while a background signal obscured the response at lower dose levels. For doses between 5 and 60 Gy, the standard deviation of single alanine film dose estimates was about 2%. The alanine film dose estimates yielded results comparable to those from the Monte Carlo simulations and the ion chamber measurements, with absolute differences between estimates in the order of 1-15%. The treatment planning calculations exhibited limited applicability. The current work shows that alanine film dosimetry is a method suitable for estimating radiotherapeutical doses and for dose measurements in inhomogeneous media.

摘要

在当前工作中,使用丙氨酸薄膜(厚度为134微米)的电子顺磁共振(EPR)剂量测定法被用于测量接受放射治疗光子束照射的非均匀体模中的剂量。主要体模材料是聚甲基丙烯酸甲酯(PMMA),而聚苯乙烯泡沫塑料或铝被用作不均匀性材料。用6或15兆伏光子将体模照射至最大剂量约30戈瑞。对丙氨酸薄膜剂量计的性能进行了研究,并与电离室剂量测定法、蒙特卡罗模拟和放射治疗治疗计划计算的结果进行了比较。结果发现,丙氨酸薄膜剂量计在约5戈瑞以上具有线性剂量响应,而在较低剂量水平时背景信号会掩盖响应。对于5至60戈瑞之间的剂量,单个丙氨酸薄膜剂量估计值的标准偏差约为2%。丙氨酸薄膜剂量估计值产生的结果与蒙特卡罗模拟和电离室测量的结果相当,估计值之间的绝对差异约为1%至15%。治疗计划计算的适用性有限。当前工作表明,丙氨酸薄膜剂量测定法是一种适用于估计放射治疗剂量和在非均匀介质中进行剂量测量的方法。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验