Al-Karmi Anan M, Ayaz Ali Asghar H, Al-Enezi Mamdouh S, Abdel-Rahman Wamied, Dwaikat Nidal
Department of Physics, King Fahd University of Petroleum & Minerals, Dhahran, 31261, Saudi Arabia.
Department of Radiation Oncology, King Fahad Specialist Hospital - Dammam, Dammam, Saudi Arabia.
Australas Phys Eng Sci Med. 2015 Sep;38(3):425-34. doi: 10.1007/s13246-015-0360-7. Epub 2015 Jul 3.
Alanine dosimeters in the form of pure alanine powder in PMMA plastic tubes were investigated for dosimetry in a clinical application. Electron paramagnetic resonance (EPR) spectroscopy was used to measure absorbed radiation doses by detection of signals from radicals generated in irradiated alanine. The measurements were performed for low-dose ranges typical for single-fraction doses often used in external photon beam radiotherapy. First, the dosimeters were irradiated in a solid water phantom to establish calibration curves in the dose range from 0.3 to 3 Gy for 6 and 18 MV X-ray beams from a clinical linear accelerator. Next, the dosimeters were placed at various locations in an anthropomorphic pelvic phantom to measure the dose delivery of a conventional four-field box technique treatment plan to the pelvis. Finally, the doses measured with alanine dosimeters were compared against the doses calculated with a commercial treatment planning system (TPS). The results showed that the alanine dosimeters have a highly sensitive dose response with good linearity and no energy dependence in the dose range and photon beams used in this work. Also, a fairly good agreement was found between the in-phantom dose measurements with alanine dosimeters and the TPS dose calculations. The mean value of the ratios of measured to calculated dose values was found to be near unity. The measured points in the in-field region passed dose-difference acceptance criterion of 3% and those in the penumbral region passed distance-to-agreement acceptance criterion of 3 mm. These findings suggest that the pure alanine powder in PMMA tube dosimeter is a suitable option for dosimetry of radiotherapy photon beams.
对以聚甲基丙烯酸甲酯(PMMA)塑料管中的纯丙氨酸粉末形式存在的丙氨酸剂量计进行了临床应用中的剂量测定研究。电子顺磁共振(EPR)光谱法用于通过检测辐照丙氨酸中产生的自由基信号来测量吸收的辐射剂量。测量针对外照射光子束放疗中常用的单分次剂量典型的低剂量范围进行。首先,将剂量计在固体水模体中进行辐照,以建立临床直线加速器6和18 MV X射线束在0.3至3 Gy剂量范围内的校准曲线。接下来,将剂量计放置在人体盆腔模体的不同位置,以测量传统四野盒式技术治疗计划对盆腔的剂量输送。最后,将用丙氨酸剂量计测量的剂量与用商业治疗计划系统(TPS)计算的剂量进行比较。结果表明,在本工作中使用的剂量范围和光子束条件下,丙氨酸剂量计具有高度敏感的剂量响应、良好的线性且无能量依赖性。此外,发现用丙氨酸剂量计进行的模体内剂量测量与TPS剂量计算之间有相当好的一致性。测量剂量值与计算剂量值之比的平均值接近1。野内区域的测量点通过了3%的剂量差异接受标准,半影区域的测量点通过了3 mm的距离一致性接受标准。这些发现表明,PMMA管剂量计中的纯丙氨酸粉末是放疗光子束剂量测定的合适选择。