Suppr超能文献

Development of reformative surgery method using partial freezing for the liver.

作者信息

Takahashi M, Nomura S, Jindai M, Shibata S, Zhu X, Watanabe Y, Kawachi K, Okabe N

出版信息

J Biomech Eng. 2006 Dec;128(6):862-6. doi: 10.1115/1.2244577.

Abstract

To minimize surgical stresses including blood loss and operation time to the patients during hepatic resection, we studied the feasibility of a combination of a partial liver freezing technique and shape-memory alloy, which also enables a free-designed resection curve. In this surgical procedure, the region surrounding a tumor in the liver is frozen to excise and prevent hemorrhage. The liver was frozen by a Peltier module. The effects of cooling rate and freezing temperature on the excision force that arise between a scalpel and the liver are carried out experimentally as a basic research for partial freezing surgical procedures. A porcine liver was used as a liver sample. The physical properties were estimated by using the finite element method based on the heat transfer characteristics of the liver. Isolation of the liver was conducted using a scalpel attached to the end-effector of a 3 degrees of freedom robot. In the experiments, the minimum excision force was obtained at a temperature between 272 K and 275 K; therefore, it is preferable that the liver be excised within this temperature range. Lowering of the cooling rate decreases the excision force even if the temperature of the liver remains unchanged. The lower the temperature of the liver is, the larger the increment rate of excision force is with regard to the cooling rate.