Geselschap Jim H, van Zuiden Jorrit M, Toonder Irwin M, Wittens Cees H A
Department of Surgery, Sint Franciscus Gasthuis, Klieweg 500, 3045 PM Rotterdam, The Netherlands.
J Endovasc Ther. 2006 Dec;13(6):762-9. doi: 10.1583/04-1208R.1.
To report the results of in vitro hydrostatic bench testing of a new vein-stent combination to correct deep venous incompetence.
Twelve valves were constructed from a modified Palmaz stent encased in a segment of great saphenous vein harvested from patients during routine varicose vein surgery. An in vitro flow circuit was set up to evaluate opening and closing pressures (in cm H(2)O), and the valve was subsequently subjected to repetitive cycles of increasing prograde flow and reflux pressures. Duplex scanning was used to evaluate valve closure time and detect any possible reflux.
The valve mechanism required only 1 to 3 cm H(2)O for opening and 2 to 4 cm H(2)O for closing. Prograde flow of up to 1000 mL/min passed easily through this valve, which remained competent with reflux pressures up to 180 cm H(2)O. Mean valve closure time was 0.15+/-0.07 seconds. Unligated side branches or damage from dissection impaired the competence of 3 valves.
This new vein-stent valve functions as a normal deep vein valve, requiring only minimal pressures for opening and closing. It allows high flow passage and still remains competent at high reflux pressures. This valve may provide a minimally invasive solution for the correction of deep venous incompetence using autologous material. Further in vivo evaluation will be mandatory.
报告一种用于纠正深静脉功能不全的新型静脉支架组合的体外流体静压试验结果。
从常规静脉曲张手术患者身上获取的一段大隐静脉包裹在改良的帕尔马兹支架内,构建了12个瓣膜。建立体外血流回路以评估开启和关闭压力(以厘米水柱计),随后对瓣膜施加递增的正向血流和反流压力的重复循环。使用双功扫描评估瓣膜关闭时间并检测任何可能的反流。
瓣膜机制开启仅需1至3厘米水柱,关闭需2至4厘米水柱。高达1000毫升/分钟的正向血流可轻松通过该瓣膜,在高达180厘米水柱的反流压力下仍保持功能正常。平均瓣膜关闭时间为0.15±0.07秒。未结扎的侧支或剥离造成的损伤损害了3个瓣膜的功能。